Line 73: | Line 73: | ||
The consistency check for beam size calculation in Q7 shows:\\ | The consistency check for beam size calculation in Q7 shows:\\ | ||
X_calculated = 3.13mm, X_cosy_read_out = 3.16mm at the Q7 exit.\\ | X_calculated = 3.13mm, X_cosy_read_out = 3.16mm at the Q7 exit.\\ | ||
- | Emittance calculated from fit parameters ε= √(ac) / L< | + | Emittance calculated from fit parameters ε= √(ac) / L< |
- | Start beam emittance= XX * AX= 0.2 π mm mrad, so the emittance grows from the start to the focal point.\\ | + | Start beam emittance= XX · AX= 0.2 π mm mrad, so the emittance grows from the start to the focal point.\\ |
This emittance grow is possibly due to the optics aberrations. That is because the input COSY file does a 4th-order-calculation. Also the measured points don't fit the parabola ideally, i.e. the optics is non-linear.\\ | This emittance grow is possibly due to the optics aberrations. That is because the input COSY file does a 4th-order-calculation. Also the measured points don't fit the parabola ideally, i.e. the optics is non-linear.\\ | ||