Line 28: | Line 28: | ||
- | Preparation of Gas bottles | + | ===== Preparation of Gas bottles |
+ | |||
+ | ==== In Data-U6 ==== | ||
+ | * Ensure I265GV is closed to isolate focal plane vacuum chamber from the rest of the beamline | ||
+ | * Confirm that the focal plane chamber is under vacuum | ||
+ | * Confirm that the only open gas handling system valves are 4, 5, 6, 20, 21, and 22 | ||
+ | |||
+ | ==== In the S3 vault ==== | ||
+ | * Ensure that the gas handling system roughing pump is running and has a good vacuum | ||
+ | * The pump sits on the floor under the focal plane behind the south support post | ||
+ | * The display unit for the pressure at the pump has a needle indicator and is mounted under the focal plane on the south support post | ||
+ | * Ensure that the main valves on the gas bottles are open | ||
+ | * Ensure that the gas bottles are not empty | ||
+ | * The gas regulators are not normally adjusted. | ||
+ | * The bottle supplying the P-10 gas for the ion chamber is secured to the cross-brace of the south support for the focal plane chamber. | ||
+ | * The bottle supplying CF4 (tetrafluoromethane) gas for the CRDCs is secured to the cross-brace of the south support for the focal plane chamber. | ||
+ | * The bottle supplying isobutane (C4H10) for the CRDCs is secured to the south support post for the focal plane chamber. | ||
- | In Data-U6 | ||
- | Ensure I265GV is closed to isolate focal plane vacuum chamber from the rest of the beamline | ||
- | Confirm that the focal plane chamber is under vacuum | ||
- | Confirm that the only open gas handling system valves are 4, 5, 6, 20, 21, and 22 | ||
- | In the S3 vault | ||
- | Ensure that the gas handling system roughing pump is running and has a good vacuum | ||
- | The pump sits on the floor under the focal plane behind the south support post (PICTURE GOES HERE) | ||
- | The display unit for the pressure at the pump has a needle indicator and is mounted under the focal plane on the south support post (PICTURE GOES HERE) | ||
- | Ensure that the main valves on the gas bottles are open | ||
- | Ensure that the gas bottles are not empty | ||
- | The gas regulators are not normally adjusted. | ||
- | The bottle supplying the P-10 gas for the ion chamber is secured to the cross-brace of the south support for the focal plane chamber. | ||
- | The bottle supplying CF4 (tetrafluoromethane) gas for the CRDCs is secured to the cross-brace of the south support for the focal plane chamber. | ||
- | The bottle supplying isobutane (C4H10) for the CRDCs is secured to the south support post for the focal plane chamber. | ||
Filling Ion Chamber with Gas | Filling Ion Chamber with Gas |