This shows you the differences between two versions of the page.
Both sides previous revision Previous revision | Next revision Both sides next revision | ||
gas_handling_system [2013/12/13 21:06] pereira |
gas_handling_system [2013/12/13 21:16] pereira |
||
---|---|---|---|
Line 33: | Line 33: | ||
* Ensure I265GV is closed to isolate focal plane vacuum chamber from the rest of the beamline | * Ensure I265GV is closed to isolate focal plane vacuum chamber from the rest of the beamline | ||
* Confirm that the focal plane chamber is under vacuum | * Confirm that the focal plane chamber is under vacuum | ||
- | * Confirm that the only open gas handling system valves are 4, 5, 6, 20, 21, and 22 | + | * Confirm that the only open gas handling system valves are **4, 5, 6, 20, 21, and 22** |
==== In the S3 vault ==== | ==== In the S3 vault ==== | ||
Line 47: | Line 47: | ||
+ | ===== Filling Ion Chamber with Gas ===== | ||
- | Filling Ion Chamber with Gas | + | * Follow instructions under [[# |
+ | * Monitor the pressure of the focal plane vacuum chamber (as read out via the I264IG ion gauge) while detectors are being filled with gas; the pressure normally stays below 5E-5 Torr. | ||
- | Follow instructions under PREPARATION FOR FILLING DETECTORS WITH GAS above | + | * Confirm that the detector pressure as indicated by the display labeled P.T. #1 reads below a couple of Torr |
- | Monitor the pressure of the focal plane vacuum chamber (as read out via the I264IG ion gauge) while detectors are being filled with gas; the pressure normally stays below 5E-5 Torr. | + | |
- | Confirm that the detector pressure as indicated by the display labeled P.T. #1 reads below a couple of Torr | + | |
- | To zero this display: | + | |
- | Click on the Slot 3 (PT1) tab to access the details page for PT #1 | + | |
- | Click the Zero Sensor button | + | |
- | Click the Valves tab at the top of the LabView window to return to the overview display of the gas handling system | + | |
- | Close valve 6 | + | |
- | Open valve 2; the flow reading at MFC #1 should jump momentarily | + | |
- | Fill Ion chamber | + | |
- | Enter the desired fill pressure and start the automatic filling process | + | |
- | Click on the tab labeled Slot 8 (M-card) IC to access the page with the pressure set point control | + | |
- | Enter the desired pressure (typically 300 Torr) in the field under the knob labeled Set Point | + | |
- | Click button labeled Return to AUTO Mode on left to start filling/ | + | |
- | Click the Valves tab at the top of the LabView window to return to the overview display of the gas handling system to monitor filling | + | |
- | The flow readings for MFC #1 will increase and eventually disappear as the readout value overflows the display field (this is normal behavior) | + | |
- | If necessary, the filling process can be stopped at any time by pressing the valve button labeled Close MFC #1 | + | |
- | Open valve 9 | + | |
- | The filling process will take several minutes | + | |
- | When the detector is filled, close valve 5 | + | |
+ | * Close **valve 6** | ||
+ | |||
+ | * Open **valve 2**; the flow reading at MFC #1 should jump momentarily | ||
+ | |||
+ | * Fill Ion chamber | ||
+ | * Enter the desired fill pressure and start the automatic filling process | ||
+ | * Click on the tab labeled **Slot 8 (M-card) IC** to access the page with the pressure set point control | ||
+ | * Enter the desired pressure (typically 300 Torr) in the field under the knob labeled Set Point | ||
+ | * Click button labeled **Return to AUTO Mode** on left to start filling/ | ||
+ | * Click the **Valves** tab at the top of the LabView window to return to the overview display of the gas handling system to monitor filling. The flow readings for MFC #1 will increase and eventually disappear as the readout value overflows the display field (this is normal behavior) | ||
+ | * If necessary, the filling process can be stopped at any time by pressing the valve button labeled Close MFC #1 | ||
+ | * Open **valve 9** | ||
+ | * The filling process will take several minutes | ||
+ | |||
+ | * When the detector is filled, close **valve 5** | ||
Filling CRDC1 and CRDC2 with Gas | Filling CRDC1 and CRDC2 with Gas |