==== In Data-U6 ==== | ==== In Data-U6 ==== | ||
- | * Ensure **I265GV** is closed (in Panel Mate S800vac.MT2 page 07) to isolate focal plane vacuum chamber from the rest of the beamline | + | * Ensure **I265GV** is closed (in PanelMate **S800vac.MT2 page 07**) to isolate focal plane vacuum chamber from the rest of the beamline |
* Confirm that the focal plane chamber is under vacuum | * Confirm that the focal plane chamber is under vacuum | ||
* Confirm that the only open gas handling system valves are **4, 5, 6, 20, 21, and 22** | * Confirm that the only open gas handling system valves are **4, 5, 6, 20, 21, and 22** | ||
Line 51: | Line 51: | ||
* Follow instructions under [[# | * Follow instructions under [[# | ||
- | * Monitor the pressure of the focal plane vacuum chamber (as read out via the I264IG ion gauge) while detectors are being filled with gas; the pressure normally stays below 5E-5 Torr. | + | * Monitor the pressure of the focal plane vacuum chamber (as read out via the I264IG ion gauge in PanelMate **S800vac.MT2 page 07**) while detectors are being filled with gas; the pressure normally stays below 5E-5 Torr. |
* Confirm that the detector pressure as indicated by the display labeled P.T. #1 reads below a couple of Torr | * Confirm that the detector pressure as indicated by the display labeled P.T. #1 reads below a couple of Torr | ||
Line 81: | Line 81: | ||
* Follow instructions under [[# | * Follow instructions under [[# | ||
- | * Monitor the pressure of the focal plane vacuum chamber (as read out via the I264IG ion gauge) while detectors are being filled with gas; the pressure normally stays below 5E-5 Torr. | + | * Monitor the pressure of the focal plane vacuum chamber (as read out via the I264IG ion gauge in PanelMate **S800vac.MT2 page 07**) while detectors are being filled with gas; the pressure normally stays below 5E-5 Torr. |
* Confirm that the detector pressure as indicated by the display labeled P.T. #2 reads below a couple of Torr | * Confirm that the detector pressure as indicated by the display labeled P.T. #2 reads below a couple of Torr |