gas_handling_system [2013/12/14 16:16] pereira [Filling Ion Chamber with Gas] |
gas_handling_system [2013/12/26 14:43] pereira [In the S3 vault] |
||
---|---|---|---|
Line 6: | Line 6: | ||
* PanelMate **S800vac.MT2 page 07**: Focal Plane Detector | * PanelMate **S800vac.MT2 page 07**: Focal Plane Detector | ||
- | The gas handling systems for both Focal Plane gas-filled detectors – the [[Detectors# | + | The gas handling systems for both Focal Plane gas-filled detectors – the [[Detectors# |
Note that since the same roughing pump drives the gas handling systems for both gas-filled detectors, extra precautions must be taken to protect one detector from the pressure of the other detector. The fact that the same pump drives both systems is not indicated on the graphics for the LabView control application | Note that since the same roughing pump drives the gas handling systems for both gas-filled detectors, extra precautions must be taken to protect one detector from the pressure of the other detector. The fact that the same pump drives both systems is not indicated on the graphics for the LabView control application | ||
Line 46: | Line 46: | ||
* The bottle supplying isobutane (C< | * The bottle supplying isobutane (C< | ||
+ | {{: | ||
+ | {{: | ||
===== Filling Ion Chamber with Gas ===== | ===== Filling Ion Chamber with Gas ===== | ||
Line 81: | Line 83: | ||
* Follow instructions under [[# | * Follow instructions under [[# | ||
- | * Monitor the pressure of the focal plane vacuum chamber (as read out via the I264IG ion gauge) while detectors are being filled with gas; the pressure normally stays below 5E-5 Torr. | + | * Monitor the pressure of the focal plane vacuum chamber (as read out via the I264IG ion gauge in PanelMate **S800vac.MT2 page 07**) while detectors are being filled with gas; the pressure normally stays below 5E-5 Torr. |
* Confirm that the detector pressure as indicated by the display labeled P.T. #2 reads below a couple of Torr | * Confirm that the detector pressure as indicated by the display labeled P.T. #2 reads below a couple of Torr | ||
Line 113: | Line 115: | ||
===== Removing Gas from Focal Plane Detectors ===== | ===== Removing Gas from Focal Plane Detectors ===== | ||
- | * Monitor the pressure of the focal plane vacuum chamber (as read out via the I264IG ion gauge) while gas is being removed from detectors; the pressure normally stays below 5E-5 Torr | + | * Monitor the pressure of the focal plane vacuum chamber (as read out via the I264IG ion gauge in PanelMate **S800vac.MT2 page 07**) while gas is being removed from detectors; the pressure normally stays below 5E-5 Torr |
* Ensure that the system is in the normal state for flowing gas through the detectors | * Ensure that the system is in the normal state for flowing gas through the detectors |