This shows you the differences between two versions of the page.
Both sides previous revision Previous revision Next revision | Previous revision Next revision Both sides next revision | ||
gas_handling_system [2013/12/14 16:18] pereira [Removing Gas from Focal Plane Detectors] |
gas_handling_system [2013/12/26 14:42] pereira [In the S3 vault] |
||
---|---|---|---|
Line 6: | Line 6: | ||
* PanelMate **S800vac.MT2 page 07**: Focal Plane Detector | * PanelMate **S800vac.MT2 page 07**: Focal Plane Detector | ||
- | The gas handling systems for both Focal Plane gas-filled detectors – the [[Detectors# | + | The gas handling systems for both Focal Plane gas-filled detectors – the [[Detectors# |
Note that since the same roughing pump drives the gas handling systems for both gas-filled detectors, extra precautions must be taken to protect one detector from the pressure of the other detector. The fact that the same pump drives both systems is not indicated on the graphics for the LabView control application | Note that since the same roughing pump drives the gas handling systems for both gas-filled detectors, extra precautions must be taken to protect one detector from the pressure of the other detector. The fact that the same pump drives both systems is not indicated on the graphics for the LabView control application | ||
Line 46: | Line 46: | ||
* The bottle supplying isobutane (C< | * The bottle supplying isobutane (C< | ||
+ | {{: | ||
+ | {{ : | ||
===== Filling Ion Chamber with Gas ===== | ===== Filling Ion Chamber with Gas ===== |