Line 34: | Line 34: | ||
* Confirm that the focal plane chamber is under vacuum | * Confirm that the focal plane chamber is under vacuum | ||
* Confirm that the only open gas handling system valves are **4, 5, 6, 20, 21, and 22** | * Confirm that the only open gas handling system valves are **4, 5, 6, 20, 21, and 22** | ||
+ | * (Check section on [[vacuum control|vacuum control]] for further details.) | ||
==== In the S3 vault ==== | ==== In the S3 vault ==== | ||
* Ensure that the gas handling system roughing pump is running and has a good vacuum | * Ensure that the gas handling system roughing pump is running and has a good vacuum | ||
* The pump sits on the floor under the focal plane behind the south support post | * The pump sits on the floor under the focal plane behind the south support post | ||
- | * The display unit for the pressure at the pump has a needle indicator and is mounted under the focal plane on the south support post | + | * The display unit for the pressure at the pump has a needle indicator and is mounted under the [[Stations# |
- | * Ensure that the main valves on the gas bottles are open | + | * Ensure that the main valves on the gas bottles are open (see pictures below) |
* Ensure that the gas bottles are not empty | * Ensure that the gas bottles are not empty | ||
- | * The gas regulators are not normally adjusted. They should all be set to provide a pressure around 20 psi. | + | * The gas regulators are not normally adjusted. They should all be set to provide a pressure around 20 psi |
- | * The bottle supplying the P-10 gas for the ion chamber is secured to the cross-brace of the south support for the focal plane chamber. | + | * The bottle supplying the P-10 gas for the [[Detectors# |
- | * The bottle supplying CF< | + | * The bottle supplying CF< |
- | * The bottle supplying isobutane (C< | + | * The bottle supplying isobutane (C< |
{{: | {{: | ||
Line 83: | Line 84: | ||
* Follow instructions under [[# | * Follow instructions under [[# | ||
- | * Monitor the pressure of the focal plane vacuum chamber (as read out via the I264IG ion gauge in PanelMate **S800vac.MT2 page 07**) while detectors are being filled with gas; the pressure normally stays below 5E-5 Torr. | + | * Monitor the pressure of the focal plane vacuum chamber (as read out via the I264IG ion gauge in PanelMate **S800vac.MT2 page 07**) while detectors are being filled with gas; the pressure normally stays below 5 x 10< |
* Confirm that the detector pressure as indicated by the display labeled P.T. #2 reads below a couple of Torr | * Confirm that the detector pressure as indicated by the display labeled P.T. #2 reads below a couple of Torr | ||
Line 147: | Line 148: | ||
* Open **valves 6 and 22** | * Open **valves 6 and 22** | ||
- | * Enter vault and close main valves on gas bottles: | + | * Enter S3 vault and close main valves on [[# |