 Required controls and monitoring Required controls and monitoring
   * **LabView gas handling system** application   * **LabView gas handling system** application
-  * PanelMate **S800vac.MT2 page 07**: Focal Plane Detector+  * PanelMate **S800vac.ui page 07**: Focal Plane Detector
  
 The gas handling systems for both Focal Plane gas-filled detectors – the [[Detectors#Ionization Chamber|Ionization Chamber]] and the [[Detectors#Cathode Readout Drift Chambers (CRDC)|CRDCs]] – are remotely controlled via a single LabView application running on the computer [[Software|S800FP-PC]] in the S3 vault. This computer can be accessed remotely via the [[Software|u6pc5]] computer in data-U6. The gas handling systems for both Focal Plane gas-filled detectors – the [[Detectors#Ionization Chamber|Ionization Chamber]] and the [[Detectors#Cathode Readout Drift Chambers (CRDC)|CRDCs]] – are remotely controlled via a single LabView application running on the computer [[Software|S800FP-PC]] in the S3 vault. This computer can be accessed remotely via the [[Software|u6pc5]] computer in data-U6.
Line 25: Line 25:
   * Once both detectors are empty, this application can be closed with the “X” button on the upper right corner of the window.   * Once both detectors are empty, this application can be closed with the “X” button on the upper right corner of the window.
  
-{{:wiki:GasHandlingSystem-panel.jpg?800|Labview control panel of the gas handling system}} +{{:wiki:gas-handling-labview.png?800|Labview control panel of the gas handling system}} 
  
  
Line 31: Line 31:
  
 ==== In Data-U6 ==== ==== In Data-U6 ====
-      * Ensure **I265GV** is closed (in PanelMate **S800vac.MT2 page 07**) to isolate focal plane vacuum chamber from the rest of the beamline+      * Ensure **I265GV** is closed (in PanelMate **S800vac.ui page 07**) to isolate focal plane vacuum chamber from the rest of the beamline
       * Confirm that the focal plane chamber is under vacuum       * Confirm that the focal plane chamber is under vacuum
       * Confirm that the only open gas handling system valves are **4, 5, 6, 20, 21, and 22**       * Confirm that the only open gas handling system valves are **4, 5, 6, 20, 21, and 22**
Line 148: Line 148:
   * Open **valves 6 and 22**   * Open **valves 6 and 22**
  
-  * Enter S3 vault and close main valves on [[#Preparation of gas bottles|gas bottles]]:  P-10 for Ion Chamber isobutane (C<sub>4</sub>H<sub>10</sub>) and CF<sub>4</sub> (tetrafluoromethane) for CRDCs+  * Enter S3 vault and close main valves on [[#Preparation of gas bottles|gas bottles]]: P-10 for the [[Detectors#Ionization Chamber|ionization chamber]] (C<sub>4</sub>H<sub>10</sub>) and CF<sub>4</sub> (tetrafluoromethane) for the [[Detectors#Cathode Readout Drift Chambers (CRDC)|CRDCs]] 
  
  