Required controls and monitoring | Required controls and monitoring | ||
* **LabView gas handling system** application | * **LabView gas handling system** application | ||
- | * PanelMate **S800vac.MT2 page 07**: Focal Plane Detector | + | * PanelMate **S800vac.ui page 07**: Focal Plane Detector |
The gas handling systems for both Focal Plane gas-filled detectors – the [[Detectors# | The gas handling systems for both Focal Plane gas-filled detectors – the [[Detectors# | ||
Line 25: | Line 25: | ||
* Once both detectors are empty, this application can be closed with the “X” button on the upper right corner of the window. | * Once both detectors are empty, this application can be closed with the “X” button on the upper right corner of the window. | ||
- | {{:wiki:GasHandlingSystem-panel.jpg? | + | {{:wiki:gas-handling-labview.png? |
Line 31: | Line 31: | ||
==== In Data-U6 ==== | ==== In Data-U6 ==== | ||
- | * Ensure **I265GV** is closed (in PanelMate **S800vac.MT2 page 07**) to isolate focal plane vacuum chamber from the rest of the beamline | + | * Ensure **I265GV** is closed (in PanelMate **S800vac.ui page 07**) to isolate focal plane vacuum chamber from the rest of the beamline |
* Confirm that the focal plane chamber is under vacuum | * Confirm that the focal plane chamber is under vacuum | ||
* Confirm that the only open gas handling system valves are **4, 5, 6, 20, 21, and 22** | * Confirm that the only open gas handling system valves are **4, 5, 6, 20, 21, and 22** | ||
Line 54: | Line 54: | ||
* Follow instructions under [[# | * Follow instructions under [[# | ||
- | * Monitor the pressure of the focal plane vacuum chamber (as read out via the I264IG ion gauge in PanelMate **S800vac.MT2 page 07**) while detectors are being filled with gas; the pressure normally stays below 5E-5 Torr. | + | * Monitor the pressure of the focal plane vacuum chamber (as read out via the I264IG ion gauge in PanelMate **S800vac.ui page 07**) while detectors are being filled with gas; the pressure normally stays below 5E-5 Torr. |
* Confirm that the detector pressure as indicated by the display labeled P.T. #1 reads below a couple of Torr | * Confirm that the detector pressure as indicated by the display labeled P.T. #1 reads below a couple of Torr | ||
Line 150: | Line 150: | ||
* Enter S3 vault and close main valves on [[# | * Enter S3 vault and close main valves on [[# | ||
+ | ===== Tracking Gas bottles pressures ===== | ||
+ | In order to track the status of the gas bottles for the [[Detectors# | ||
+ | {{: |