Line 31: Line 31:
  
 ==== In Data-U6 ==== ==== In Data-U6 ====
-      * Ensure **I265GV** is closed (in PanelMate **S800vac.MT2 page 07**) to isolate focal plane vacuum chamber from the rest of the beamline+      * Ensure **I265GV** is closed (in PanelMate **S800vac.ui page 07**) to isolate focal plane vacuum chamber from the rest of the beamline
       * Confirm that the focal plane chamber is under vacuum       * Confirm that the focal plane chamber is under vacuum
       * Confirm that the only open gas handling system valves are **4, 5, 6, 20, 21, and 22**       * Confirm that the only open gas handling system valves are **4, 5, 6, 20, 21, and 22**
Line 54: Line 54:
   * Follow instructions under [[#Preparation of Gas bottles|Preparation of Gas bottles]]    * Follow instructions under [[#Preparation of Gas bottles|Preparation of Gas bottles]] 
  
-  * Monitor the pressure of the focal plane vacuum chamber (as read out via the I264IG ion gauge in PanelMate **S800vac.MT2 page 07**) while detectors are being filled with gas; the pressure normally stays below 5E-5 Torr.+  * Monitor the pressure of the focal plane vacuum chamber (as read out via the I264IG ion gauge in PanelMate **S800vac.ui page 07**) while detectors are being filled with gas; the pressure normally stays below 5E-5 Torr.
  
   * Confirm that the detector pressure as indicated by the display labeled P.T. #1 reads below a couple of Torr   * Confirm that the detector pressure as indicated by the display labeled P.T. #1 reads below a couple of Torr
Line 150: Line 150:
   * Enter S3 vault and close main valves on [[#Preparation of gas bottles|gas bottles]]: P-10 for the [[Detectors#Ionization Chamber|ionization chamber]] (C<sub>4</sub>H<sub>10</sub>) and CF<sub>4</sub> (tetrafluoromethane) for the [[Detectors#Cathode Readout Drift Chambers (CRDC)|CRDCs]]    * Enter S3 vault and close main valves on [[#Preparation of gas bottles|gas bottles]]: P-10 for the [[Detectors#Ionization Chamber|ionization chamber]] (C<sub>4</sub>H<sub>10</sub>) and CF<sub>4</sub> (tetrafluoromethane) for the [[Detectors#Cathode Readout Drift Chambers (CRDC)|CRDCs]] 
  
 +===== Tracking Gas bottles pressures =====
 +In order to track the status of the gas bottles for the ion chamber and the CRDCs, it is possible to monitor their remaining pressures from Barney. Just open the **A1900** page, then **Tools**, and then **Gas Status**. The following window will be displayed:
  
  
  
  
 +[[Detectors#Ionization Chamber|ionization chamber]] 
 +[[Detectors#Cathode Readout Drift Chambers (CRDC)|CRDCs]] 
 +[[Detectors#[Tracking Parallel Plate Avalanche Counters (TPPAC)]|TPPAC]] 