====== Detector gas handling system ====== | ====== Detector gas handling system ====== | ||
+ | * [[#General information]] | ||
+ | * [[#LabView Control Program]] | ||
+ | * [[# | ||
+ | * [[#Filling Ion Chamber with Gas]] | ||
+ | * [[#Filling CRDC1 and CRDC2 with Gas]] | ||
+ | * [[#Removing Gas from Focal Plane Detectors]] | ||
+ | * [[#Tracking Gas bottles pressures]] | ||
+ | |||
===== General Information ===== | ===== General Information ===== | ||
Required controls and monitoring | Required controls and monitoring | ||
- | | + | * LabView gas handling system |
* PanelMate **S800vac.ui page 07**: Focal Plane Detector | * PanelMate **S800vac.ui page 07**: Focal Plane Detector | ||
- | The gas handling systems for both Focal Plane gas-filled detectors – the [[Detectors# | + | The gas handling systems for both Focal Plane gas-filled detectors – the [[Detectors# |
Note that since the same roughing pump drives the gas handling systems for both gas-filled detectors, extra precautions must be taken to protect one detector from the pressure of the other detector. The fact that the same pump drives both systems is not indicated on the graphics for the LabView control application | Note that since the same roughing pump drives the gas handling systems for both gas-filled detectors, extra precautions must be taken to protect one detector from the pressure of the other detector. The fact that the same pump drives both systems is not indicated on the graphics for the LabView control application | ||
Line 155: | Line 163: | ||
- | {{: | + | {{: |