gas_handling_system

Differences

This shows you the differences between two versions of the page.

Link to this comparison view

Both sides previous revision Previous revision
Next revision
Previous revision
Next revision Both sides next revision
gas_handling_system [2017/01/22 16:47]
pereira [Detector gas handling system]
gas_handling_system [2017/05/14 10:04]
pereira [Filling Ion Chamber with Gas]
Line 2: Line 2:
  
   * [[#General information]]   * [[#General information]]
- +  * [[#LabView Control Program]] 
 +  * [[#Preparation of Gas bottles|Preparation of Gas bottles for filling of detectors]] 
 +  * [[#Filling Ion Chamber with Gas]] 
 +  * [[#Filling CRDC1 and CRDC2 with Gas]] 
 +  * [[#Removing Gas from Focal Plane Detectors]] 
 +  * [[#Tracking Gas bottles pressures|Tracking Gas bottles pressures from Barney]] 
 +  * [[#Replacing Gas bottles]] 
 ===== General Information ===== ===== General Information =====
 Required controls and monitoring Required controls and monitoring
Line 72: Line 79:
           * Click on the tab labeled **Slot 8 (M-card) IC** to access the page with the pressure set point control           * Click on the tab labeled **Slot 8 (M-card) IC** to access the page with the pressure set point control
           * Enter the desired pressure (typically 300 Torr) in the field under the knob labeled Set Point           * Enter the desired pressure (typically 300 Torr) in the field under the knob labeled Set Point
 +          * Click button labeled **Update Set Point**
           * Click button labeled **Return to AUTO Mode** on left to start filling/regulation           * Click button labeled **Return to AUTO Mode** on left to start filling/regulation
           * Click the **Valves** tab at the top of the LabView window to return to the overview display of the gas handling system to monitor filling. The flow readings for MFC #1 will increase and eventually disappear as the readout value overflows the display field (this is normal behavior)           * Click the **Valves** tab at the top of the LabView window to return to the overview display of the gas handling system to monitor filling. The flow readings for MFC #1 will increase and eventually disappear as the readout value overflows the display field (this is normal behavior)
-      * If necessary, the filling process can be stopped at any time by pressing the valve button labeled **Close MFC #1** 
       * Open **valve 9**       * Open **valve 9**
 +      * If necessary, the filling process can be stopped at any time by pressing the valve button labeled **Close MFC #1**
       * The filling process will take several minutes       * The filling process will take several minutes
  
Line 107: Line 115:
           * Click on the tab labeled **Virtual Ch. 0 (CRDC)** to access the page with the pressure set point control           * Click on the tab labeled **Virtual Ch. 0 (CRDC)** to access the page with the pressure set point control
           * Enter the desired pressure (typically 40 Torr) in the field under the knob labeled **Set Point**           * Enter the desired pressure (typically 40 Torr) in the field under the knob labeled **Set Point**
-          * Click button labeled **Return to AUTO Mode** on left to start filling/regulation+          * Click button labeled **AUTO Mode** on left to start filling/regulation
           * Click the **Valves** tab at the top of the LabView window to return to the overview display of the gas handling system to monitor filling. The flow readings for MFC #2 and MFC #3 will increase. The LabView application controls the gas mixing ratio based on the flow ratio (scaled with a calibration constant)           * Click the **Valves** tab at the top of the LabView window to return to the overview display of the gas handling system to monitor filling. The flow readings for MFC #2 and MFC #3 will increase. The LabView application controls the gas mixing ratio based on the flow ratio (scaled with a calibration constant)
 +      * Open **valve 25**
       * If necessary, the filling process can be stopped at any time by pressing the valve button labeled **Close MFCs #2 & #3**       * If necessary, the filling process can be stopped at any time by pressing the valve button labeled **Close MFCs #2 & #3**
-      * Open **valve 25** 
       * The filling process will take several minutes       * The filling process will take several minutes
  
Line 158: Line 166:
  
 {{:wiki:gas-bottles-status.jpg?900|Barney display of pressures of all "supported" gas detectors in the lab}} {{:wiki:gas-bottles-status.jpg?900|Barney display of pressures of all "supported" gas detectors in the lab}}
 +
 +
 +
 +
 +===== Replacing Gas bottles =====
 +In preparation