- | {{:wiki:hv-gui.png?700|GUI control | + | The high-voltage of the plastic scintillator and the MCP at Object, the [[Detectors# |
+ | |||
+ | {{:wiki:phoebus_diag_hv.png|CS-Studio Phoebus DIAG HV control}} | ||
+ | |||
+ | This page can be reached from “FRIB EXP” → “DIAG” (Engineering Area, not to be confused with “S800” → “Diag Main”) → “S800 HV Bias Power Supply”. | ||
+ | |||
+ | The ramp rate can be changed in the " | ||
+ | |||
+ | Each channel can also be accessed from by going to “FRIB EXP” -> “Power Supplies” and open the “EXP-S3” tab, which looks like the screenshot below. | ||
+ | |||
+ | {{: | ||
+ | The actual HV modules are located on the top level of the S3 vault, under the FP chamber. | ||
- | ===== HV Manual control ==== | + | {{: |
- | Apart from the detectors listed above, the voltages of the [[Detectors# | + | |
- | and the **electrostatic gate** of the [[Detectors# | + | |
- | The bias for the object scintillator is controlled via the H.V. power supply {{: | + | Other examples |
- | In order for the [[Detectors# | + | * **CRDC1 and CRDC2**: |
- | + | * For He-3 @ ~130 MeV/u: CRDC1 (Anode=1120 V, Drift=1000 V); CRDC2 (Anode=1120 V; Drift=1000 V) | |
+ | * For Be-12 @ ~30 MeV/u: CRDC1 (Anode=970 V, Drift=1000 V); CRDC2 (Anode=970 V; Drift=1000 V) | ||
+ | * For Ar-34 @ ~80 MeV/u: CRDC1 (Anode=860 V, Drift=1000 V); CRDC2 (Anode=840 V; Drift=1000 V) | ||
+ | * For Zn-58 @ ~44 MeV/u: CRDC1 (Anode=740 V, Drift=1000 V); CRDC2 (Anode=740 V; Drift=1000 V) | ||
+ | * For Rb-74 @ ~40 MeV/u: CRDC1 (Anode=650 | ||
+ | * For U-238 (88+) @ ~70 MeV/u: CRDC1 (Anode=570 V, Drift=1000 V); CRDC2 (Anode=570 V; Drift=1000 V) | ||
+ | |||
+ | * **FP scintillator (E1 up and E1 down)**: | ||
+ | * For He-3 @ ~130 MeV/u: UP (1790 V); DOWN (1760 V) | ||
+ | * For Be-12 @ ~30 MeV/u: UP (1770 V); DOWN (1700 V) | ||
+ | * For Ar-34 @ ~80 MeV/u: UP (1580 V); DOWN (1510 V) | ||
+ | * For Zn-58 @ ~44 MeV/u: UP (1400 V); DOWN (1340 V) | ||
+ | * For U-238 (88+) @ ~70 MeV/u: UP (1180 V); DOWN (1280 V) | ||
- | {{:wiki: | + | * **IC**: |
+ | * Typically 200 V | ||