installation_fp_sci [2015/05/04 12:24]
pereira [Open focal-plane chamber]
installation_fp_sci [2015/05/04 12:48]
pereira [Installation of FP scintillator]
Line 13: Line 13:
   - After venting the FP chamber, there might still be some pressure differencial. This can be corrected by "forcing" the chamber flange {{:wiki:screwdriver-flange.jpg?350}}   - After venting the FP chamber, there might still be some pressure differencial. This can be corrected by "forcing" the chamber flange {{:wiki:screwdriver-flange.jpg?350}}
   - Unlock the padlock protecting the power brakers of the switch controls to open the chamber, and press the **START** button {{:wiki:padlock-note.jpg?350|Padlock}}   - Unlock the padlock protecting the power brakers of the switch controls to open the chamber, and press the **START** button {{:wiki:padlock-note.jpg?350|Padlock}}
-  - Turn the switch to open the clamps holding the FP chamber The weight held by the pivot will increase {{:wiki:switch-clamps.jpg?250|Switch clamps}} {{:wiki:clamps.jpg?350|Clamps holding FP chamber}} {{:wiki:scale.jpg?350|Scale FP chamber}}+  - Turn the switch to open the clamps holding the FP chamberThe weight held by the hydraulic piston will increase {{:wiki:switch-clamps.jpg?250|Switch clamps}} {{:wiki:clamps.jpg?400|Clamps holding FP chamber}} {{:wiki:scale.jpg?350|Scale FP chamber}} 
 +  - Turn on the switch to activate the motor opening the chamber. It takes about 10 minutes to get the chamber totally opened {{:wiki:fp-chamber-motor.jpg?250|Motor switch to open FP chamber}}
  
 +
 +==== Installation of FP scintillator ====
 +  - The figure below shows the FP chamber open, with the "old" FP scintillator to be replaced by the new one {{:wiki:E1-view.jpg?350|FP chamber opened}}
 +  - Unplug the bases of the up and down phototubes. The platic tie holding the cable of the up phototube to the chamber needs to be cut
 +{{:wiki:e1-up-Daniel.jpg?400|UP base phototube}}   {{:wiki:e1-down-closeup.jpg?250|DOWN base phototube}}
 +  - Using an allen wrench, unscrew the two bolts attaching the scintillator frame to the base of the chamber {{:wiki:e1-frame-bolts.jpg?350|Scintillator frame}}
 +
 +he first step to do is unplugging the up and down bases of the phototubes 
 +
 +
 +In order to remove the "old" scintillator, we need first to u
  
  
 Click Vent status [start pumpdown] --> vent status= pumpdown in progress (blue) Click Vent status [start pumpdown] --> vent status= pumpdown in progress (blue)