Line 17: Line 17:
  
  
-==== Installation of FP scintillator ====+==== Closing focal-plane chamber ==== 
   - The figure below shows the FP chamber open, with the "old" FP scintillator to be replaced by the new one {{:wiki:E1-view.jpg?350|FP chamber opened}}   - The figure below shows the FP chamber open, with the "old" FP scintillator to be replaced by the new one {{:wiki:E1-view.jpg?350|FP chamber opened}}
   - Unplug the bases of the up and down phototubes. The platic tie holding the cable of the up phototube to the chamber needs to be cut {{:wiki:e1-up-Daniel.jpg?400|UP base phototube}}   {{:wiki:e1-down-closeup.jpg?250|DOWN base phototube}}   - Unplug the bases of the up and down phototubes. The platic tie holding the cable of the up phototube to the chamber needs to be cut {{:wiki:e1-up-Daniel.jpg?400|UP base phototube}}   {{:wiki:e1-down-closeup.jpg?250|DOWN base phototube}}
Line 35: Line 36:
  
 Click Vent status [start pumpdown] --> vent status= pumpdown in progress (blue) Click Vent status [start pumpdown] --> vent status= pumpdown in progress (blue)
 +
 +
 +
 +
 +
 +  - Turn the switch to move the chamber to position "UP", and keep it that way for about 10 minutes (the time it takes the chamber to be closed)
 +  - When the chamber is almost close, turn the know to close the clamps
 +  - Before pumping the chamber down, make sure that the scintillator is properly connected. This can be done by biasing the detector and checking the signals, either from the scalers or with a scope. One should clearly see cosmic rays; if that is not the case, the phototubes are likely miss connected
 +  - In order to pumpdown the chamber, make sure that the manual vent valve is closed
 +  - Start the manual pump to get a pressure below 50 mtorr {{:wiki:ManualValve-fp.jpg?350|Manual valve pump}}
 +  - 
 +  - 