==== Vent focal-plane chamber ==== | ==== Vent focal-plane chamber ==== | ||
- | | + | |
- | | + | |
- | | + | |
- | | + | |
- | | + | * Open PanelMate and select **S800vac** file in directory \\exp\files\control\ncsdata\General\Qt\Templates. Go to page 07, click on the **VENT STATUS** window and begin the venting procedure. The **VENT VALVE I260VV** window will turn red and display " |
- | | + | |
+ | {{: | ||
+ | | ||
==== Open focal-plane chamber ==== | ==== Open focal-plane chamber ==== | ||
Line 17: | Line 19: | ||
- | ==== Closing focal-plane chamber | + | ==== Installing FP scintillator |
- | - Turn the switch to move the chamber to position " | + | |
- | - When the chamber is almost close, turn the know to close the clamps | + | |
- | - Before pumping the chamber down, make sure that the scintillator is properly connected. This can be done by biasing the detector and checking the signals, either from the scalers or with a scope. One should clearly see cosmic rays; if that is not the case, the phototubes are likely miss connected | + | |
- | - In order to pumpdown the chamber, make sure that the manual vent valve is closed | + | |
- | - Start the manual pump to get a pressure below 50 mtorr {{: | + | |
- | - | + | |
- | - | + | |
- The figure below shows the FP chamber open, with the " | - The figure below shows the FP chamber open, with the " | ||
- Unplug the bases of the up and down phototubes. The platic tie holding the cable of the up phototube to the chamber needs to be cut {{: | - Unplug the bases of the up and down phototubes. The platic tie holding the cable of the up phototube to the chamber needs to be cut {{: | ||
Line 32: | Line 28: | ||
- Connect the bases of the up and down phototubes. The later is quite difficult because one cannot see the pins | - Connect the bases of the up and down phototubes. The later is quite difficult because one cannot see the pins | ||
- Using a plastic tie, attach the cable of the up phototube to the scintillator frame | - Using a plastic tie, attach the cable of the up phototube to the scintillator frame | ||
- | - | ||
- | he first step to do is unplugging the up and down bases of the phototubes | ||
- | In order to remove the " | ||
- | + | ==== Closing focal-plane chamber | |
- | Click Vent status [start pumpdown] | + | - Turn the switch to move the chamber to position " |
+ | - When the chamber is almost close, turn the know to close the clamps | ||
+ | - Before pumping the chamber down, make sure that the scintillator is properly connected. This can be done by biasing the detector and checking the signals, either from the scalers or with a scope. One should clearly see cosmic rays; if that is not the case, the phototubes are likely miss connected | ||
+ | - In order to pumpdown the chamber, make sure that the manual vent valve is closed. Start manual pump to get a pressure below 50 mtorr {{: | ||
+ | - Close the manual pump; open PanelMate **S800vac** and go to page 07. Select window **PUMPDOWN STATUS** and click on button **START PUMPDOWN**. The **PUMPDOWN STATUS** and **VENT STATUS** windows should display " | ||
+ | - Wait for some hours until vacuum is good |