installation_fp_sci [2015/05/04 13:11] pereira [Closing focal-plane chamber] |
installation_fp_sci [2017/03/06 12:19] pereira [Vent focal-plane chamber] |
||
---|---|---|---|
Line 4: | Line 4: | ||
- Open the [[hv_bias|HV Remote Control]] GUI and ensure that the focal-plane detectors are unbiased | - Open the [[hv_bias|HV Remote Control]] GUI and ensure that the focal-plane detectors are unbiased | ||
- Open the [[gas_handling_system# | - Open the [[gas_handling_system# | ||
- | - Ensure that the manual venting valve located on the south side of the chamber is closed (see figures below) | + | - Ensure that the manual venting valve located on the south side of the chamber is closed (see figures below) |
+ | - Ensure that the FP ion gauge (I264IG) is off (see details [[vacuum_control|here]]) | ||
- Open PanelMate and select **S800vac** file in directory \\exp\files\control\ncsdata\General\Qt\Templates. Go to page 07, click on the **VENT STATUS** window and begin the venting procedure. The **VENT VALVE I260VV** window will turn red and display " | - Open PanelMate and select **S800vac** file in directory \\exp\files\control\ncsdata\General\Qt\Templates. Go to page 07, click on the **VENT STATUS** window and begin the venting procedure. The **VENT VALVE I260VV** window will turn red and display " | ||
- | - Gently open the manual venting valve until the pressure displayed in the Pirani Gauge (see figure below) increases by ~4 torr/min {{: | + | - Gently open the manual venting valve until the pressure displayed in the Pirani Gauge (see figure below) increases by ~1 torr every 15 seconds |
- Wait until the chamber is vented (few hours). The **I261PG** Pirani Gauge should read atmospheric pressure. The window **VENT STATUS** in PanelMate should display "VENT VALVE OPEN". The window **PUMPDOWN STATUS** should display " | - Wait until the chamber is vented (few hours). The **I261PG** Pirani Gauge should read atmospheric pressure. The window **VENT STATUS** in PanelMate should display "VENT VALVE OPEN". The window **PUMPDOWN STATUS** should display " | ||