====== Installation of focal-plane scintillator ====== | ====== Installation of focal-plane scintillator ====== | ||
- | ==== Vent focal-plane chamber ==== | + | ===== Vent focal-plane chamber |
* Open the [[hv_bias|HV Remote Control]] GUI and ensure that the focal-plane detectors are unbiased | * Open the [[hv_bias|HV Remote Control]] GUI and ensure that the focal-plane detectors are unbiased | ||
* Open the [[gas_handling_system# | * Open the [[gas_handling_system# | ||
Line 11: | Line 11: | ||
* Wait until the chamber is vented (few hours). The **I261PG** Pirani Gauge should read slightly above atmospheric pressure (due to the dry N2 flow pumped to the chamber). The window **VENT STATUS** in PanelMate should display "VENT VALVE OPEN". The window **PUMPDOWN STATUS** should display " | * Wait until the chamber is vented (few hours). The **I261PG** Pirani Gauge should read slightly above atmospheric pressure (due to the dry N2 flow pumped to the chamber). The window **VENT STATUS** in PanelMate should display "VENT VALVE OPEN". The window **PUMPDOWN STATUS** should display " | ||
- | ==== Open focal-plane chamber ==== | + | ===== Open focal-plane chamber |
- | * After venting the FP chamber, the pressure inside the chamber is slightly above atmospheric due to the dry N2 pumped. Before opening the chamber, it is recommended to vent the extra N2 by " | + | * After venting the FP chamber, the pressure inside the chamber is slightly above atmospheric due to the dry N2 pumped. Before opening the chamber, it is recommended to vent the extra N2 by " |
* Unlock the padlock protecting the power brakers of the switch controls to open the chamber, and press the **START** button {{: | * Unlock the padlock protecting the power brakers of the switch controls to open the chamber, and press the **START** button {{: | ||
* Turn the switch to open the clamps holding the FP chamber. The weight held by the hydraulic piston will increase {{: | * Turn the switch to open the clamps holding the FP chamber. The weight held by the hydraulic piston will increase {{: | ||
Line 19: | Line 19: | ||
- | ==== Installing FP scintillator ==== | + | ===== Installing FP scintillator |
* The figure below shows the FP chamber open, with the " | * The figure below shows the FP chamber open, with the " | ||
Line 33: | Line 33: | ||
- | ==== Closing focal-plane chamber ==== | + | ===== Closing focal-plane chamber |
- | | + | |
- | | + | |
- | | + | |
- | - In order to pumpdown the chamber, make sure that the manual vent valve is closed. Start manual pump to get a pressure below 50 mtorr {{: | + | |
- | | + | ===== Pumping focal-plane chamber ===== |
- | | + | |
+ | * In order to pumpdown the chamber, make sure that the manual vent valve is closed. Start manual pump to get a pressure below 50 mtorr {{: | ||
+ | | ||
+ | |