====== Installation of focal-plane scintillator ====== | ====== Installation of focal-plane scintillator ====== | ||
+ | In June 2019, a new Turbo Pump with magnetically-levitating bearings (Leybold MAG 2200iP) was installed in the S800 focal-plane chamber. This pump has a dry-N line connected to it that is purging continuously the bearings. This creates an extra pressure in the foreline, which must be pumping down continuously. Therefore, it is critical turn off the turbo pump whenever the foreline valve needs to be closed. | ||
+ | |||
+ | Instructions on how to vent/ | ||
===== Vent focal-plane chamber ===== | ===== Vent focal-plane chamber ===== | ||
- | * Open the [[hv_bias|HV Remote Control]] GUI and ensure that the focal-plane detectors are unbiased | + | * Open the [[hv_bias|HV Remote Control]] GUI and ensure that the gas focal-plane detectors |
- | * Open the [[gas_handling_system# | + | * Make sure that the CRDC grid gate is off (HV is in dU6) |
+ | * Make sure that the CRDC preamps and interface boards are turned off (it is located in the NIM crate attached to the S800 focal-plane chamber) | ||
+ | * In order to avoid condensation from the cooling system, turn off the chiller sitting under the S800 focal-plane chamber. Wait for at least a couple of hours before venting the chamber. | ||
+ | * Open the [[gas_handling_system# | ||
* Ensure that the manual venting valve located on the south side of the chamber is closed (see figures below) | * Ensure that the manual venting valve located on the south side of the chamber is closed (see figures below) | ||
* Ensure that the FP ion gauge (I264IG) is off (see details [[vacuum_control|here]]) | * Ensure that the FP ion gauge (I264IG) is off (see details [[vacuum_control|here]]) | ||
- | * Open PanelMate and select **S800vac** file in directory \\exp\files\control\ncsdata\General\Qt\Templates. Go to page 07, and ensure that I265GV (gate valve separating FP box from rest of beam line) is closed. | + | * Open PanelMate and select **S800vac** file in directory \\exp\files\control\ncsdata\General\Qt\Templates, and go to page 07 (see figure below) |
- | * Click on the **VENT STATUS** window and begin the venting procedure. The **VENT VALVE I260VV** window will turn red and display " | + | {{: |
+ | * Close I265GV (gate valve separating FP box from rest of beam line) is closed. | ||
+ | * Click on the **VENT STATUS** window and begin the venting procedure. The **VENT VALVE I260VV** window will turn red and display " | ||
* Gently open the manual venting valve until the pressure displayed in the Pirani Gauge (see figure below) increases by ~1 torr every 15 seconds (see figures below) | * Gently open the manual venting valve until the pressure displayed in the Pirani Gauge (see figure below) increases by ~1 torr every 15 seconds (see figures below) | ||
{{: | {{: | ||
+ | * Make sure that pressure inside the CRDCs and IC detectors (LabView) increase at the same rate than the pressure in the chamber. Otherwise, stop the venting process inmediately. | ||
* Wait until the chamber is vented (few hours). The **I261PG** Pirani Gauge should read slightly above atmospheric pressure (due to the dry N2 flow pumped to the chamber). The window **VENT STATUS** in PanelMate should display "VENT VALVE OPEN". The window **PUMPDOWN STATUS** should display " | * Wait until the chamber is vented (few hours). The **I261PG** Pirani Gauge should read slightly above atmospheric pressure (due to the dry N2 flow pumped to the chamber). The window **VENT STATUS** in PanelMate should display "VENT VALVE OPEN". The window **PUMPDOWN STATUS** should display " | ||
- | * When the chamber is fully vented, close the manual vent valve | + | |
+ | * When the Turbo pump is fully stopped, switch off the chiller located near the S800 electronics rack | ||
+ | | ||
+ | * Close the manual vent valve | ||
===== Open focal-plane chamber ===== | ===== Open focal-plane chamber ===== | ||
Line 37: | Line 49: | ||
===== Closing focal-plane chamber ===== | ===== Closing focal-plane chamber ===== | ||
* Turn the switch to move the chamber to position " | * Turn the switch to move the chamber to position " | ||
- | * When the chamber is almost | + | * Check the display indicating the weight lift by the hydraulic system. |
+ | * Turn the knob to close the clamps | ||
+ | * Inspect the clamps and chamber flange (no gaps) to ensure that the chamber is properly sealed | ||
* Before pumping the chamber down, make sure that the scintillator is properly connected. This can be done by biasing the detector and checking the signals, either from the scalers or with a scope. One should clearly see cosmic rays; if that is not the case, the phototubes are likely miss connected | * Before pumping the chamber down, make sure that the scintillator is properly connected. This can be done by biasing the detector and checking the signals, either from the scalers or with a scope. One should clearly see cosmic rays; if that is not the case, the phototubes are likely miss connected | ||
===== Pumping focal-plane chamber ===== | ===== Pumping focal-plane chamber ===== | ||
* Make sure that the pressure-relief valve and the manual vent valve (see above) are closed. | * Make sure that the pressure-relief valve and the manual vent valve (see above) are closed. | ||
- | * In order to pumpdown the chamber, start manual pump to get a pressure below 50 mtorr {{: | + | * Make sure that manual |
- | * Close the manual pump; open PanelMate **S800vac** and go to page 07. Select window **PUMPDOWN STATUS** and click on button **START PUMPDOWN**. The **PUMPDOWN STATUS** and **VENT STATUS** windows should display " | + | * Open the gas-handling-system LabView control and make sure that the only open detector valves are **4** (IC supply), **5** (IC bypass), **6** (IC focal-plane bypass), **7** (Ion Chamber vent), **20** (CRDC supply), **21** (CRDC bypass), and **22** (CRDC focal-plane bypass) and **23** (CRDC vent). The rest of the valves should be closed. The pressure inside the detectors should be ~760 Torr. |
- | * Wait for some hours until vacuum is good | + | * Open PanelMate **S800vac** and go to page 07. Make sure that the venting valve **I260VV** is closed (see last step of the [[#Vent focal-plane chamber|venting process]] described above) |
+ | * Make sure that the turbo pump is off (as seen in PanelMate **s800vac**, | ||
+ | * Select window **PUMPDOWN STATUS** and click on button **START PUMPDOWN**. The **PUMPDOWN STATUS** and **VENT STATUS** windows should display " | ||
+ | * The vent valve **I260VV** and foreline valve **I260FV** will be automatically closed | ||
+ | * After the rough valve **I260RV** is automatically open, the mechanical pump will start pumping. | ||
+ | * The pressures in the mechanical pump (from **I259PG**) and beam chamber (from **I261PG**) will jump momentarily to very high values. They should go down slowly after few seconds (otherwise, the chamber is not properly closed). | ||
+ | * Make sure that pressure inside the CRDCs and IC (LabView) decrease at the same rate as the chamber. Otherwise, stop the pump-down sequence immediately. | ||
+ | * Note that during the rough pump-down sequence, pressure in the foreline I260PG will increase to few torr due to the purge line connected to the turbo pump. This is ok, as long as the turbo pump is OFF | ||
+ | * When pressure in **I261PG** is ~95 mTorr, turn on the chiller providing cooling water to the turbo pump. | ||
+ | * When pressure in the chamber (from **I261PG**) gets below 90 mTorr (at least 45 min), **I260RV** will close. | ||
+ | * Open the foreline valve **I260FV** using PanelMate box **Foreline vlve I260FV ** and clicking " | ||
+ | * When pressure in the foreline (**I260PG**) is below 200 mTorr, turn on the turbo pump. You will see the speed increasing in the PanelMate box **I260TP speed** | ||
+ | * It takes about 8 min to get the turbo pump "At Speed" (as read in box **Turbo pump focal plane detector I260TP**). During that time, the pressure in the focal-plane chamber will slowly increase (it is not being pump down) by few mTorr | ||
+ | * When the turbo pump is "At Speed" and pressure in the chamber (as read in **I261PG**) is below 175 mTorr, the turbo-pump gate valve **I260GV** will open | ||
+ | * After **I260GV** gets opened, you will see a pressure spike in the foreline **I260PG** up to ~1 Torr. This is normal behavior. This pressure will slowly decrease below 200 mTorr in about 2 minutes. It will take at least 30 min to reach the nominal value (~60 mTorr). | ||
+ | * Wait for few hours until vacuum is good. You can then turn on the ion gauge **I264IG** | ||
+ |