Line 16: Line 16:
   - Click **Enable Controls** for the CRDC and IC Control Valve PIDs and then click **DISABLE** for each of them. This will open the control valve to 100% for each vessel and begin pumping them out.   - Click **Enable Controls** for the CRDC and IC Control Valve PIDs and then click **DISABLE** for each of them. This will open the control valve to 100% for each vessel and begin pumping them out.
   - The IC vessel has a much larger quantity of gas compared to the CRDC vessel. In order to pump them out at a similar speed go to the “FP_State_Buttons” page and go to the D2708 (IC vessel) tab. Click **MAN** to enable manual mode. Click **OPN** for “S800_GHS:SV_D2708:OPN_CMD_10” to open an additional valve to the pump.   - The IC vessel has a much larger quantity of gas compared to the CRDC vessel. In order to pump them out at a similar speed go to the “FP_State_Buttons” page and go to the D2708 (IC vessel) tab. Click **MAN** to enable manual mode. Click **OPN** for “S800_GHS:SV_D2708:OPN_CMD_10” to open an additional valve to the pump.
-  - Open a data browser and plot the PVs for PT1 and PT2. +  - Open a data browser and plot the PVs for PT1 (S800_GHS:CMG_D2708A:VP_RD) and PT2 (S800_GHS:CMG_D2703A:VP_RD)
   - Wait for PT1 and PT2 to read < 2 Torr. This should take about 30 minutes.   - Wait for PT1 and PT2 to read < 2 Torr. This should take about 30 minutes.
  
 ===== Venting FP Chamber ===== ===== Venting FP Chamber =====
   - Ensure the IG in the Focal Plane Pumping Station is off. This page can be found in FRIB EXP Main → S800 → Vacuum Main (Primary Pages) → Focal Plane.   - Ensure the IG in the Focal Plane Pumping Station is off. This page can be found in FRIB EXP Main → S800 → Vacuum Main (Primary Pages) → Focal Plane.
-  - Return to the “FP_State_Buttons” page and under the D2708 tab click **AUTO**. This will close all valves and allow the system to enter Vent mode+  - Return to the “FP_State_Buttons” page and for each of the D2708 and D2703 tabs click **PUMP**.  
-  - Click **VENT**. This will expose the vessels to the Pumping Station. Observe the FPG on the “D2430 Pumping Station” page to ensure the Turbo Pump is not overloaded.+  - Click **VENT** for each tab. This will expose the vessels to the Pumping Station. Observe the FPG on the “D2686 Pumping Station” page to ensure the Turbo Pump is not overloaded.
   - Once the FPG bottoms out, check the box next to “Valve Controls”, then click **Close** next to the Turbo Gate Valve (TGV).    - Once the FPG bottoms out, check the box next to “Valve Controls”, then click **Close** next to the Turbo Gate Valve (TGV). 
   - If the system will only be vented for a short time, leave the TMP on, otherwise turn it off.   - If the system will only be vented for a short time, leave the TMP on, otherwise turn it off.
-  - Plot PT1, PT2, and PT3 from the “GHS_FP_Main” page as well as BPG from the “D2430 Pumping Station” page on the same axis.+  - Plot PT1 (S800_GHS:CMG_D2708A:VP_RD), PT2 (S800_GHS:CMG_D2703A:VP_RD), and PT3 (S800_GHS:CMG_D2708C:VP_RD) from the “GHS_FP_Main” page as well as BPG (RBT_BTS34:BPG_D2686:VP_RD) from the “D2686 Pumping Station” page on the same axis.
   - Briefly (~2 seconds) open the vent valve by clicking **Open** next to the valve labeled “Vent” then click **Close**. Check that all the pressures are increasing at the same rate.   - Briefly (~2 seconds) open the vent valve by clicking **Open** next to the valve labeled “Vent” then click **Close**. Check that all the pressures are increasing at the same rate.
   - Again, open the vent valve and observe the pressure for 30 seconds to ensure the pressures are increasing at the same rate among all four gauges. If a difference of > 1 Torr is observed between any two gauges, click **Close** on the vent valve.   - Again, open the vent valve and observe the pressure for 30 seconds to ensure the pressures are increasing at the same rate among all four gauges. If a difference of > 1 Torr is observed between any two gauges, click **Close** on the vent valve.
Line 38: Line 38:
   - Open “Focal Plane Pumping Station” page from FRIB EXP Main → S800 → Vacuum Main (Primary Pages).   - Open “Focal Plane Pumping Station” page from FRIB EXP Main → S800 → Vacuum Main (Primary Pages).
   - Open “GHS_FP_Main” and “FP_State_Buttons” pages from FRIB EXP Main → S800 → GHS Pages (Secondary Pages).   - Open “GHS_FP_Main” and “FP_State_Buttons” pages from FRIB EXP Main → S800 → GHS Pages (Secondary Pages).
-  - Plot PT1, PT2, and PT3 from the “GHS_FP_Main” page as well as BPG from the “D2430 Pumping Station” page on the same axis.+  - Plot PT1 (S800_GHS:CMG_D2708A:VP_RD), PT2 (S800_GHS:CMG_D2703A:VP_RD), and PT3 (S800_GHS:CMG_D2708C:VP_RD) from the “GHS_FP_Main” page as well as BPG (RBT_BTS34:BPG_D2686:VP_RD) from the “D2686 Pumping Station” page on the same axis.
   - On the “FP_State_Buttons” page change from “Vent” mode to “Pump” mode by clicking **PUMP** in both the D2703 and D2708 tabs.   - On the “FP_State_Buttons” page change from “Vent” mode to “Pump” mode by clicking **PUMP** in both the D2703 and D2708 tabs.
   - Check the box next to “Valve Controls” on the “D2686 Pumping Station” page.   - Check the box next to “Valve Controls” on the “D2686 Pumping Station” page.
-  - If the TMP is off and Foreline Valve (FV) is closed, open the Roughing Valve (RV) briefly (2 seconds) by clicking **Open** and observe the behavior of the pressure gauges. If they all decrease at the same rate, again click **Open** to begin rouging out the chambers. +  - If the TMP is off and Foreline Valve (FV) is closed, open the Roughing Valve (RV) briefly (2 seconds) by clicking **Open** and observe the behavior of the pressure gauges. If they all decrease at the same rate, again click **Open** to begin roughing out the chambers. 
-  - If the TMP is still on (this could be the case for a short vent period), click **Close** on the FV to protect the pump before doing step 6. +  - If the TMP is still on (this could be the case for a short vent period), turn it off and close the foreline valve (FVbefore doing step 6. 
-  - When the BPG reads zero, **Close** RV and **Open** FV. Turn the TMP on if it is not already. When the TMP is “at speed” **Open** the TGV. If the TMP was off before opening the TGV, the chamber pressure may have risen to a point where the TGV closes automatically. If the BPG still reads < 1 Torrattempt to **Open** the TGV again.+  - When the BPG reads below 200 mTorr, **Close** RV and **Open** FV. Turn the TMP on if it is not already. When the TMP is “at speed” (~30,000 RPM), **Open** the TGV.  
 +  - NOTE: before TMP reaches "at speed"pressure in the chamber, as read by BPG, might increase to a point where TGV can not be open. If that is the caseturn off the turbo pump, close FV, and repeat steps 6, 7, 8 but this time, wait for BPG to read below 100 mTorr. Then try again step 8.
   - Allow the TMP to pump on the system until the IG reads < 1E-5 Torr.   - Allow the TMP to pump on the system until the IG reads < 1E-5 Torr.
   - On the “FP_State_Buttons” page click **RUN** to enter run mode.   - On the “FP_State_Buttons” page click **RUN** to enter run mode.
Line 49: Line 50:
 ===== Filling & Running FP Gaseous Detectors ===== ===== Filling & Running FP Gaseous Detectors =====
  
-  - Plot PT1, PT2, and the control valves for both D2703 (CRDC) and D2708 (IC).+  - Plot PT1 (S800_GHS:CMG_D2708A:VP_RD), PT2 (S800_GHS:CMG_D2703A:VP_RD), and the control valves for both D2703 (CRDC, S800_GHS:CV_D2703:POS_RD_VLV) and D2708 (IC, S800_GHS:CV_D2708:POS_RD_VLV).
   - Click **Enable Controls** for the CRDC and IC “Control Valve PID”s, then click **ENABLE** to turn on PID controls for each vessel. Each control valve should adjust to its BASE value.   - Click **Enable Controls** for the CRDC and IC “Control Valve PID”s, then click **ENABLE** to turn on PID controls for each vessel. Each control valve should adjust to its BASE value.
   - If the CRDC valve remains completely open (stuck at 100%), open a probe with PV: “S800_GHS:CV_D2703:MODE_CSET_PVC”. In the “New Value” box, enter “CLOSE” and press enter. If the valve then closes (100% to 0%), enter “AUTO” in the “New Value” box and press enter. This should reenable PID control.   - If the CRDC valve remains completely open (stuck at 100%), open a probe with PV: “S800_GHS:CV_D2703:MODE_CSET_PVC”. In the “New Value” box, enter “CLOSE” and press enter. If the valve then closes (100% to 0%), enter “AUTO” in the “New Value” box and press enter. This should reenable PID control.