   - Open “Focal Plane Pumping Station” page from FRIB EXP Main → S800 → Vacuum Main (Primary Pages).   - Open “Focal Plane Pumping Station” page from FRIB EXP Main → S800 → Vacuum Main (Primary Pages).
   - Open “GHS_FP_Main” and “FP_State_Buttons” pages from FRIB EXP Main → S800 → GHS Pages (Secondary Pages).   - Open “GHS_FP_Main” and “FP_State_Buttons” pages from FRIB EXP Main → S800 → GHS Pages (Secondary Pages).
-  - Plot PT1, PT2, and PT3 from the “GHS_FP_Main” page as well as BPG from the “D2430 Pumping Station” page on the same axis.+  - Plot PT1, PT2, and PT3 from the “GHS_FP_Main” page as well as BPG from the “D2686 Pumping Station” page on the same axis.
   - On the “FP_State_Buttons” page change from “Vent” mode to “Pump” mode by clicking **PUMP** in both the D2703 and D2708 tabs.   - On the “FP_State_Buttons” page change from “Vent” mode to “Pump” mode by clicking **PUMP** in both the D2703 and D2708 tabs.
   - Check the box next to “Valve Controls” on the “D2686 Pumping Station” page.   - Check the box next to “Valve Controls” on the “D2686 Pumping Station” page.
   - If the TMP is off and Foreline Valve (FV) is closed, open the Roughing Valve (RV) briefly (2 seconds) by clicking **Open** and observe the behavior of the pressure gauges. If they all decrease at the same rate, again click **Open** to begin rouging out the chambers.   - If the TMP is off and Foreline Valve (FV) is closed, open the Roughing Valve (RV) briefly (2 seconds) by clicking **Open** and observe the behavior of the pressure gauges. If they all decrease at the same rate, again click **Open** to begin rouging out the chambers.
-  - If the TMP is still on (this could be the case for a short vent period), click **Close** on the FV to protect the pump before doing step 6. +  - If the TMP is still on (this could be the case for a short vent period), turn it off and close the foreline valve (FVbefore doing step 6. 
-  - When the BPG reads zero, **Close** RV and **Open** FV. Turn the TMP on if it is not already. When the TMP is “at speed” **Open** the TGV. If the TMP was off before opening the TGV, the chamber pressure may have risen to a point where the TGV closes automatically. If the BPG still reads < 1 Torrattempt to **Open** the TGV again.+  - When the BPG reads below 200 mTorr, **Close** RV and **Open** FV. Turn the TMP on if it is not already. When the TMP is “at speed” **Open** the TGV.  
 +  - NOTE: before reach "at speed"pressure in the chamber, as read by BPG, might increase to a point where TGV can not be open. If that is the caseturn off the turbo pump, close FV, and repeat steps 6, 7, 8 but this time, wait for BPG to read below 100 mTorr. Then try again step 8.
   - Allow the TMP to pump on the system until the IG reads < 1E-5 Torr.   - Allow the TMP to pump on the system until the IG reads < 1E-5 Torr.
   - On the “FP_State_Buttons” page click **RUN** to enter run mode.   - On the “FP_State_Buttons” page click **RUN** to enter run mode.