installation_fp_sci [2023/02/17 20:12] pereira [Pumpdown of FP Chamber] |
installation_fp_sci [2023/03/07 08:28] pereira [Venting FP Chamber] |
||
---|---|---|---|
Line 25: | Line 25: | ||
- Once the FPG bottoms out, check the box next to “Valve Controls”, | - Once the FPG bottoms out, check the box next to “Valve Controls”, | ||
- If the system will only be vented for a short time, leave the TMP on, otherwise turn it off. | - If the system will only be vented for a short time, leave the TMP on, otherwise turn it off. | ||
- | - Plot PT1, PT2, and PT3 from the “GHS_FP_Main” page as well as BPG from the “D2686 Pumping Station” page on the same axis. | + | - Plot PT1 (S800_GHS: |
- Briefly (~2 seconds) open the vent valve by clicking **Open** next to the valve labeled “Vent” then click **Close**. Check that all the pressures are increasing at the same rate. | - Briefly (~2 seconds) open the vent valve by clicking **Open** next to the valve labeled “Vent” then click **Close**. Check that all the pressures are increasing at the same rate. | ||
- Again, open the vent valve and observe the pressure for 30 seconds to ensure the pressures are increasing at the same rate among all four gauges. If a difference of > 1 Torr is observed between any two gauges, click **Close** on the vent valve. | - Again, open the vent valve and observe the pressure for 30 seconds to ensure the pressures are increasing at the same rate among all four gauges. If a difference of > 1 Torr is observed between any two gauges, click **Close** on the vent valve. | ||
Line 43: | Line 43: | ||
- If the TMP is off and Foreline Valve (FV) is closed, open the Roughing Valve (RV) briefly (2 seconds) by clicking **Open** and observe the behavior of the pressure gauges. If they all decrease at the same rate, again click **Open** to begin rouging out the chambers. | - If the TMP is off and Foreline Valve (FV) is closed, open the Roughing Valve (RV) briefly (2 seconds) by clicking **Open** and observe the behavior of the pressure gauges. If they all decrease at the same rate, again click **Open** to begin rouging out the chambers. | ||
- If the TMP is still on (this could be the case for a short vent period), turn it off and close the foreline valve (FV) before doing step 6. | - If the TMP is still on (this could be the case for a short vent period), turn it off and close the foreline valve (FV) before doing step 6. | ||
- | - When the BPG reads zero, **Close** RV and **Open** FV. Turn the TMP on if it is not already. When the TMP is “at speed” **Open** the TGV. If the TMP was off before | + | - When the BPG reads below 200 mTorr, **Close** RV and **Open** FV. Turn the TMP on if it is not already. When the TMP is “at speed” |
+ | - NOTE: before | ||
- Allow the TMP to pump on the system until the IG reads < 1E-5 Torr. | - Allow the TMP to pump on the system until the IG reads < 1E-5 Torr. | ||
- On the “FP_State_Buttons” page click **RUN** to enter run mode. | - On the “FP_State_Buttons” page click **RUN** to enter run mode. |