This shows you the differences between two versions of the page.
Both sides previous revision Previous revision | Next revision Both sides next revision | ||
installation_fp_sci [2023/02/17 20:45] pereira [Pumpdown of FP Chamber] |
installation_fp_sci [2023/02/17 20:45] pereira [Pumpdown of FP Chamber] |
||
---|---|---|---|
Line 44: | Line 44: | ||
- If the TMP is still on (this could be the case for a short vent period), turn it off and close the foreline valve (FV) before doing step 6. | - If the TMP is still on (this could be the case for a short vent period), turn it off and close the foreline valve (FV) before doing step 6. | ||
- When the BPG reads below 200 mTorr, **Close** RV and **Open** FV. Turn the TMP on if it is not already. When the TMP is “at speed” (~30,000 RPM), **Open** the TGV. | - When the BPG reads below 200 mTorr, **Close** RV and **Open** FV. Turn the TMP on if it is not already. When the TMP is “at speed” (~30,000 RPM), **Open** the TGV. | ||
- | - NOTE: before | + | - NOTE: before |
- Allow the TMP to pump on the system until the IG reads < 1E-5 Torr. | - Allow the TMP to pump on the system until the IG reads < 1E-5 Torr. | ||
- On the “FP_State_Buttons” page click **RUN** to enter run mode. | - On the “FP_State_Buttons” page click **RUN** to enter run mode. |