- Open “Focal Plane Pumping Station” page from FRIB EXP Main → S800 → Vacuum Main (Primary Pages). | - Open “Focal Plane Pumping Station” page from FRIB EXP Main → S800 → Vacuum Main (Primary Pages). | ||
- Open “GHS_FP_Main” and “FP_State_Buttons” pages from FRIB EXP Main → S800 → GHS Pages (Secondary Pages). | - Open “GHS_FP_Main” and “FP_State_Buttons” pages from FRIB EXP Main → S800 → GHS Pages (Secondary Pages). | ||
- | - Plot PT1, PT2, and PT3 from the “GHS_FP_Main” page as well as BPG from the “D2686 Pumping Station” page on the same axis. | + | - Plot PT1 (S800_GHS: |
- On the “FP_State_Buttons” page change from “Vent” mode to “Pump” mode by clicking **PUMP** in both the D2703 and D2708 tabs. | - On the “FP_State_Buttons” page change from “Vent” mode to “Pump” mode by clicking **PUMP** in both the D2703 and D2708 tabs. | ||
- Check the box next to “Valve Controls” on the “D2686 Pumping Station” page. | - Check the box next to “Valve Controls” on the “D2686 Pumping Station” page. | ||
Line 50: | Line 50: | ||
===== Filling & Running FP Gaseous Detectors ===== | ===== Filling & Running FP Gaseous Detectors ===== | ||
- | - Plot PT1, PT2, and the control valves for both D2703 (CRDC) and D2708 (IC). | + | - Plot PT1 (S800_GHS: |
- Click **Enable Controls** for the CRDC and IC “Control Valve PID”s, then click **ENABLE** to turn on PID controls for each vessel. Each control valve should adjust to its BASE value. | - Click **Enable Controls** for the CRDC and IC “Control Valve PID”s, then click **ENABLE** to turn on PID controls for each vessel. Each control valve should adjust to its BASE value. | ||
- If the CRDC valve remains completely open (stuck at 100%), open a probe with PV: “S800_GHS: | - If the CRDC valve remains completely open (stuck at 100%), open a probe with PV: “S800_GHS: |