Line 21: Line 21:
   - The IC vessel has a much larger quantity of gas compared to the CRDC vessel. In order to pump them out at a similar speed go to the “FP_State_Buttons” page and go to the D2708 (Ion Chamber) tab. Click **EMPTY**. This will more rapidly pump down the IC vessel. Both vessels should be emptied at about the same time.   - The IC vessel has a much larger quantity of gas compared to the CRDC vessel. In order to pump them out at a similar speed go to the “FP_State_Buttons” page and go to the D2708 (Ion Chamber) tab. Click **EMPTY**. This will more rapidly pump down the IC vessel. Both vessels should be emptied at about the same time.
   - Open a data browser and plot the PVs for PT1 (S800_GHS:CMG_D2708A:VP_RD) and PT2 (S800_GHS:CMG_D2703A:VP_RD).    - Open a data browser and plot the PVs for PT1 (S800_GHS:CMG_D2708A:VP_RD) and PT2 (S800_GHS:CMG_D2703A:VP_RD). 
-  - Wait for PT1 and PT2 to read < Torr. This should take about 30 minutes.+  - Wait for PT1 and PT2 to read < 0.5 Torr. This should take about 30 minutes.
   - Open the Focal Plane VAC D2686 page (FRIB EXP Main → S800 → Vacuum Main → Focal Plane).   - Open the Focal Plane VAC D2686 page (FRIB EXP Main → S800 → Vacuum Main → Focal Plane).
   - Ensure the beamline gate valve to the FP chamber is closed (BGV_D2671).   - Ensure the beamline gate valve to the FP chamber is closed (BGV_D2671).
Line 32: Line 32:
  
 ===== Venting FP Chamber ===== ===== Venting FP Chamber =====
 +
 +Notify the Control Room that you will vent the FP chamber.
 +
   - Ensure the IG in the Focal Plane Pumping Station is off. This page can be found in FRIB EXP Main → S800 → Vacuum Main (Primary Pages) → Focal Plane.   - Ensure the IG in the Focal Plane Pumping Station is off. This page can be found in FRIB EXP Main → S800 → Vacuum Main (Primary Pages) → Focal Plane.
   - Check the box next to “Valve Controls”, then click **Close** next to the Turbo Gate Valve (TGV).   - Check the box next to “Valve Controls”, then click **Close** next to the Turbo Gate Valve (TGV).