===== Description ===== | ===== Description ===== | ||
- | The NMR running programs for both the S800 analysis line and spectrograph dipoles can be started by clicking on the icons labeled S800A_NMR and S800S_NMR, located on the desktop | + | The NMR running programs for both the S800 analysis line and spectrograph dipoles can be started by selecting |
Line 16: | Line 16: | ||
===== Operation ===== | ===== Operation ===== | ||
- | The NMR program constantly checks the status of the various dipoles from the EPICS system and updates the status column. In addition, the program checks the log file at startup for the closest previously measured value. In case no previous measurement can be used the program searches for a signal around the guessed value of the field. This search can take some time if the calibration is off or the hysteresis of the dipole is large. | + | The NMR program constantly checks the status of the various dipoles from the EPICS system and updates the status column. In addition, the program checks the log file at startup for the closest previously measured value. In case no previous measurement can be used the program searches for a signal around the guessed value of the field. This search can take some time if the calibration is off or the hysteresis of the dipole is large. The maximum number of attempts is limited to 200. After that the status is marked as " |
In order to reduce the searching time, the program tries to guess the field from the following sources, in priority order: | In order to reduce the searching time, the program tries to guess the field from the following sources, in priority order: | ||
Line 26: | Line 26: | ||
* Calibration if no previous measurement is available | * Calibration if no previous measurement is available | ||
- | In addition, the program checks the log file at startup for the closest previously measured value. In case no previous measurement can be used the program searches for a signal around the guessed value of the field. This search can take some time if the calibration is off or the hysteresis of the dipole is large. The maximum number of attempts is limited to 200. After that the status is marked as “Failed” and the program gives up on this dipole. | ||
- | |||
Once a signal is found, the measuring sequence tries to find two resonance positions on either side of 0. Each resonance is calculated as the average between the negative and positive modulation minima (blue and green curves, and corresponding arrows). The final field is interpolated from those measurements, | Once a signal is found, the measuring sequence tries to find two resonance positions on either side of 0. Each resonance is calculated as the average between the negative and positive modulation minima (blue and green curves, and corresponding arrows). The final field is interpolated from those measurements, | ||
- | If a dipole is turned off or its field is being changed, the status shows either | + | If a dipole is turned off or its field is being changed, the status shows either |
{{: | {{: | ||
- | The magnet currents, measured fields, and deviation from the set value are logged in a file that is soft-linked from the " | + | The magnet currents, measured fields, and deviation from the set value are logged in a file that is soft-linked from the " |
Line 47: | Line 45: | ||
===== Matching ===== | ===== Matching ===== | ||
- | {{: | + | The button labeled “Match” opens a new window for matching |
- | + | ||
- | {{: | + | |
- | The button labeled “Match” opens a new window for matching | + | {{: |
===== Troubleshooting ===== | ===== Troubleshooting ===== |