Due to the location of the NMR probes in the dipoles of both the S800 analysis line and spectrograph, | Due to the location of the NMR probes in the dipoles of both the S800 analysis line and spectrograph, | ||
- | The solution | + | The solution implemented on the S800 uses digital oscilloscopes to digitize the NMR signal. The digitized signal is then used in a program that calculates the position of the resonance and deduces the value of the magnetic field. This program (written in Tcl/Tk) is fully autonomous and automatically writes the measured Brho values to the relevant EPICS channels. In addition, it offers the possibility to match the dipoles by pairs and keeps a log file of the measured fields. |
===== Description ===== | ===== Description ===== | ||
- | The NMR running programs for both the S800 analysis line and spectrograph dipoles can be started by selecting the NMR option in the operation applications of the desktop **devop1**. The picture below shows the NMR panel for the S800 Analysis Line GUI as an example. The top (yellow) panel shows the data relevant to the NMR module, the middle (pink) the digitized signal from the oscilloscope and the bottom (blue) the status and measurements for the dipoles. | + | The NMR running programs for both the S800 analysis line and spectrograph dipoles can be started by selecting the NMR option in the operation applications of the desktop **u6pc5**. The picture below shows the NMR panel for the S800 Analysis Line GUI as an example. The top (yellow) panel shows the data relevant to the NMR module, the middle (pink) the digitized signal from the oscilloscope and the bottom (blue) the status and measurements for the dipoles. |
Line 29: | Line 29: | ||
- | If a dipole is turned off or its field is being changed, the status shows either " | + | If a dipole is turned off or its field is being changed, the status shows either " |
- | + | {{:wiki:barney_s800_unmatched.jpg?850|Barney display for the S800.}} | |
- | THIS FIGURE IS OBSOLETE. NEW A NEWER VERSION | + | |
- | + | ||
- | {{:wiki:barney_magnets.jpg?700|Barney display for the S800.}} | + | |
Line 69: | Line 66: | ||
- Enter any **username** | - Enter any **username** | ||
- **set priv ** | - **set priv ** | ||
- | - The system will ask you the password (ask the S800 device physicist)** | + | - The system will ask you the password (ask the S800 device physicist) |
- **unattach port x** (where x=1 for the analysis line and x=2 for the spectrograph) | - **unattach port x** (where x=1 for the analysis line and x=2 for the spectrograph) | ||
- **logout** | - **logout** | ||
- Click again on the starting icon and the program should start. | - Click again on the starting icon and the program should start. | ||