====== NMR Control ===== | ====== NMR Control ===== | ||
- | Two digital oscilloscopes are dedicated to NMR readout – one for the analysis line NMRs and one for the spectrograph NMRs. The scopes are located in data-U6, in upper two shelves on the right electronics rack. They are isolated from clean ground because this signals from the NMR probes are on a dirty ground. | + | Two digital oscilloscopes are dedicated to NMR readout – one for the analysis line NMRs and one for the spectrograph NMRs. The scopes are located in the S3 mezzanine. They are isolated from clean ground because this signals from the NMR probes are on a dirty ground. |
===== Motivation ===== | ===== Motivation ===== | ||
Line 9: | Line 9: | ||
===== Description ===== | ===== Description ===== | ||
- | The NMR running programs for both the S800 analysis line and spectrograph dipoles can be started | + | The NMR running programs for both the S800 analysis line and spectrograph dipoles can be started |
- | Sometimes the initialization of these applications fail. In that case check section [[nmr probes# | + | $ ssh [your-user-name-on-EXP]@exp-s3-nmr -Y |
+ | |||
+ | and typing, for the analysis line dipoles | ||
+ | |||
+ | $ / | ||
+ | |||
+ | and for the spectrograph dipoles | ||
+ | |||
+ | $ / | ||
+ | |||
+ | **Note that it is not simply wish anymore but wish8.6.** | ||
+ | Now press start on the GUI, which is shown below, and you should see a field reading. | ||
{{: | {{: | ||
+ | ===== Terminating processes run by another user ===== | ||
+ | |||
+ | A solution has been implemented for the case where one user needs to run the NMR software via the ' | ||
+ | |||
+ | If a user is running ' | ||
+ | |||
+ | * Log on to exp-s3-nmr virtual machine (as described above) | ||
+ | * Type 'ps aux | grep wish8.6' | ||
+ | * Let's suppose user ' | ||
+ | * This will stop both wish8.6 processes that are running as user ' | ||
+ | * Now user ' | ||
+ | |||
+ | In short, | ||
+ | $ ps aux | grep wish8.6 | ||
+ | $ sudo -u usernameRunningWish8.6 pkill wish8.6 | ||
+ | | ||
+ | and you should be able to run the NMR software. | ||
===== Operation ===== | ===== Operation ===== | ||
The NMR program constantly checks the status of the various dipoles from the EPICS system and updates the status column. In addition, the program checks the log file at startup for the closest previously measured value. In case no previous measurement can be used the program searches for a signal around the guessed value of the field. This search can take some time if the calibration is off or the hysteresis of the dipole is large. The maximum number of attempts is limited to 200. After that the status is marked as " | The NMR program constantly checks the status of the various dipoles from the EPICS system and updates the status column. In addition, the program checks the log file at startup for the closest previously measured value. In case no previous measurement can be used the program searches for a signal around the guessed value of the field. This search can take some time if the calibration is off or the hysteresis of the dipole is large. The maximum number of attempts is limited to 200. After that the status is marked as " | ||
Line 51: | Line 79: | ||
===== Troubleshooting ===== | ===== Troubleshooting ===== | ||
+ | |||
+ | |||
+ | ==== NMR GUI does not start ==== | ||
The NMR program is connected to the following devices and systems: | The NMR program is connected to the following devices and systems: | ||
Line 72: | Line 103: | ||
- Click again on the starting icon and the program should start. | - Click again on the starting icon and the program should start. | ||
+ | |||
+ | ==== NMR GUI does not communicate with NMR controllers ==== | ||
+ | * This can be seen, for instance, when the NMR controller finds a resonance, but the NMR GUI keeps trying forever. | ||
+ | * Most likely, you need to reboot the terminal server located in S3 | ||
+ | * Go to the mezzanine level, where the NMR controllers are located | ||
+ | * The terminal server is behind the NMR controllers | ||
+ | * Simply turn it off, wait few seconds, and turn it back on | ||
+ | |||
+ | |||
+ | |||
+ | ==== NMR reading sometimes takes too long (> 30 min) ==== | ||
+ | |||
+ | This is typically caused by one of two possible factors: | ||
+ | |||
+ | 1. The signal is weak, and so the NMR " | ||
+ | * This can be alleviated by changing the NMR thresholds (it can be done for each magnet in the analysis line, or for both magnets in the spectrograph). | ||
+ | * Stop NMR; enter new threshold (in Volts) and click “Set”; start NMR. | ||
+ | * In general, a threshold 0.15 mV is good for normal NMR signals, whereas 0.10 mV may be necessary if the signal are too weak. | ||
+ | |||
+ | 2. Initial field where NMR starts scanning is far away from actual value (this has only been seen with Spectrograph) | ||
+ | * Check what field is NMR scanning and compare with field read by neighboring NMR. | ||
+ | * If it is too far, stop NMR; enter field (in T) and click “Set”; start NMR. | ||
+ | |||
+ | |||
+ | |||
+ | |||
+ | ==== Manually setting Dipole Fields | ||
+ | (Based on notes from Mathias Steiner with some extra info included) | ||
+ | |||
+ | You will need to get access to the NMR controllers in the mezzanine level of S3 (see picture below). You will also have to connect an oscilloscope to one of the NMR controllers, | ||
- | ====== Manually setting Dipole Fields ====== | ||
Let’s assume that you want to set the dipole I265DS (first dipole (D1) of S800 Spectrograph) to the rigidity (Brho) 3.7965 Tm (this works for any dipole in the S800) | Let’s assume that you want to set the dipole I265DS (first dipole (D1) of S800 Spectrograph) to the rigidity (Brho) 3.7965 Tm (this works for any dipole in the S800) | ||
Line 83: | Line 143: | ||
* Get access to the NMR controller in the mezzanine level of S3 vault (see figure below). Push the “SEARCH LOCAL” button to take control out of REMOTE mode. | * Get access to the NMR controller in the mezzanine level of S3 vault (see figure below). Push the “SEARCH LOCAL” button to take control out of REMOTE mode. | ||
* Select the right probe by turning the “probe knob” (see figure below). Remember, each dipole has two probes: probe 4, covers fields in the range 0.34 – 1.05 T, and probe 5 covers fields in the range 0.7–2.1 T. In this case, we should set the knob of the Spectrograph controller to position B | * Select the right probe by turning the “probe knob” (see figure below). Remember, each dipole has two probes: probe 4, covers fields in the range 0.34 – 1.05 T, and probe 5 covers fields in the range 0.7–2.1 T. In this case, we should set the knob of the Spectrograph controller to position B | ||
- | * The list of probes that can be selected with the knobs are, for the Spectrograph: | + | * The list of probes that can be selected with the knobs are, for the Spectrograph: |
* Use the “coarse” and “fine” knobs to select the field that you want (in T): 1.35454 T. By doing this, you are asking the NMR controller to just look at that field. | * Use the “coarse” and “fine” knobs to select the field that you want (in T): 1.35454 T. By doing this, you are asking the NMR controller to just look at that field. | ||
* If the field of that dipole happens to be close enough to the field that the NMR controller is looking at, Bingo! You’ll most likely see a resonance in the oscilloscope. In the most likely situation, however, your magnet is far away from the NMR search window and you will see only a flat line in the oscilloscope. | * If the field of that dipole happens to be close enough to the field that the NMR controller is looking at, Bingo! You’ll most likely see a resonance in the oscilloscope. In the most likely situation, however, your magnet is far away from the NMR search window and you will see only a flat line in the oscilloscope. | ||
- | * All you need to do now is to scan the magnetic field of your dipole, for instance using the NCS application with a knob box, and drive it towards the value that the NMR controller is looking at | + | * All you need to do now is to vary the magnetic field of your dipole, and drive it towards the value that the NMR controller is looking at. For instance, in the s800 computer **u6pc5**, open the QtKM file BLSetup_A1900.gkm (connected to the "knob box") and use channel I265DS. |
* As you get close to that value, the oscilloscope will show the NMR resonance. Keep moving until the resonance is well centered. At that point, your dipole is set to the desired field | * As you get close to that value, the oscilloscope will show the NMR resonance. Keep moving until the resonance is well centered. At that point, your dipole is set to the desired field | ||
- | **Manually setting a dipole (method | + | **Method |
* Alternatively, | * Alternatively, | ||
* Set EPICS channel **I265DS_MAG.BSET** to 13.5454 (note, the default value of this channel is always an (impossible) 9999.1111, which is ignored. Typing 13.5454 will cause the magnet to go to 257.577 A, and **I265DS_MAG.BSET** goes back to 9999.1111. Setting the magnet this way will get you to plus or minus half a percent of where you want to be. | * Set EPICS channel **I265DS_MAG.BSET** to 13.5454 (note, the default value of this channel is always an (impossible) 9999.1111, which is ignored. Typing 13.5454 will cause the magnet to go to 257.577 A, and **I265DS_MAG.BSET** goes back to 9999.1111. Setting the magnet this way will get you to plus or minus half a percent of where you want to be. | ||
* At that point, the quickest way may be to use the FINE knob in the NMR controller to find the actual field value. Just turn the knob until the resonance is centered in the oscilloscope, | * At that point, the quickest way may be to use the FINE knob in the NMR controller to find the actual field value. Just turn the knob until the resonance is centered in the oscilloscope, | ||
- | * Once you know the value of the field, type it (in Tesla) into the EPICS channel **I265DS_MAG.FELD**. Note that is what Daniel' | + | * Once you know the value of the field, type it (in Tesla) into the EPICS channel **I265DS_MAG.FELD**. Note that this is what Daniel' |
* Having done that, you can use the BARNEY ' | * Having done that, you can use the BARNEY ' | ||
* Repeat until satisfied. | * Repeat until satisfied. |