User Tools

Site Tools


nmr_probes

Differences

This shows you the differences between two versions of the page.

Link to this comparison view

Both sides previous revision Previous revision
Next revision
Previous revision
Next revision Both sides next revision
nmr_probes [2019/10/17 10:29]
pereira [Manually setting Dipole Fields]
nmr_probes [2023/02/13 11:23]
noji [Description] Updated to reflect recent changes as to how to run the NMR GUI.
Line 1: Line 1:
 ====== NMR Control ===== ====== NMR Control =====
-Two digital oscilloscopes are dedicated to NMR readout – one for the analysis line NMRs and one for the spectrograph NMRs. The scopes are located in data-U6, in upper two shelves on the right electronics rack. They are isolated from clean ground because this signals from the NMR probes are on a dirty ground.+Two digital oscilloscopes are dedicated to NMR readout – one for the analysis line NMRs and one for the spectrograph NMRs. The scopes are located in the S3 mezzanine. They are isolated from clean ground because this signals from the NMR probes are on a dirty ground.
  
 ===== Motivation ===== ===== Motivation =====
Line 9: Line 9:
  
 ===== Description ===== ===== Description =====
-The NMR running programs for both the S800 analysis line and spectrograph dipoles can be started from the desktop of computer [[Software#u6pc5 (data U6)|u6pc5]], by clicking in the icons **NMR Analysis** for the analysis line dipoles, and **NMR Spectrograph** for the spectrograph magnets. Alternatively, the NMR control programs can be run from a Linux session by typing ''/usr/bin/wish analysis.tcl'' in directory **/user/s800/Documents/Run/analysis**, for the analysis line, and ''/usr/bin.wish spectrograph_sjw.tcl'' in directory **/user/s800/Documents/Run/spectrograph**, for the spectrograph.+The NMR running programs for both the S800 analysis line and spectrograph dipoles can be started by logging in to exp-s3-nmr.ftc (10.40.230.111) accessible from the EXP network, namely, from a whatever computer connected to the EXP network, 
 + 
 +  $ ssh [your-user-name-on-EXP]@exp-s3-nmr -Y 
 + 
 +and typing, for the analysis line dipoles  
 + 
 +  $ /usr/bin/wish /srv/s3-nmr/analysis/analysis.tcl 
 + 
 +and for the spectrograph dipoles 
 + 
 +  $ /usr/bin/wish /srv/s3-nmr/spectrograph/spectrograph.tcl
  
 Sometimes the initialization of these applications fail. In that case check section [[nmr probes#troubleshooting|"Troubleshooting"]] below). The picture below shows the NMR panel for the S800 Analysis Line GUI as an example. The top (yellow) panel shows the data relevant to the NMR module, the middle (pink) the digitized signal from the oscilloscope and the bottom (blue) the status and measurements for the dipoles.  Sometimes the initialization of these applications fail. In that case check section [[nmr probes#troubleshooting|"Troubleshooting"]] below). The picture below shows the NMR panel for the S800 Analysis Line GUI as an example. The top (yellow) panel shows the data relevant to the NMR module, the middle (pink) the digitized signal from the oscilloscope and the bottom (blue) the status and measurements for the dipoles. 
Line 51: Line 61:
  
 ===== Troubleshooting ===== ===== Troubleshooting =====
 +
 +
 +==== NMR GUI does not start ====
  
 The NMR program is connected to the following devices and systems:  The NMR program is connected to the following devices and systems: 
Line 73: Line 86:
  
  
 +==== NMR GUI does not communicate with NMR controllers ====
 +  * This can be seen, for instance, when the NMR controller finds a resonance, but the NMR GUI keeps trying forever.
 +  * Most likely, you need to reboot the terminal server located in S3
 +  * Go to the mezzanine level, where the NMR controllers are located
 +  * The terminal server is behind the NMR controllers
 +  * Simply turn it off, wait few seconds, and turn it back on
 +
 +
 +
 +==== Spectrograph NMR GUI "locks" in noise or does not see weak resonance====
 +  * This is typically the case when the thresholds in the NMR GUI are not set properly. If they are too low, the program may lock in noise. Conversely, if the thresholds are too high, it may miss a weak resonance.
 +  * Threshold for the Spectrograph resonance can be adjusted by hand in the Spectrograph NMR GUI. In general, a threshold 0.15 mV is good for normal NMR signals, whereas 0.10 mV may be necessary if the signal are too weak.
  
  
 ==== Manually setting Dipole Fields  ==== ==== Manually setting Dipole Fields  ====
 (Based on notes from Mathias Steiner with some extra info included) (Based on notes from Mathias Steiner with some extra info included)
 +
 +You will need to get access to the NMR controllers in the mezzanine level of S3 (see picture below). You will also have to connect an oscilloscope to one of the NMR controllers, either analysis line (left) or Spectrograph (right). Connect the left BNC connector (field modulation) to CH1 in the oscilloscope and the right BNC connector (NMR signal) in CH2. Select XY display, and AC coupling.
 +
 +
  
 Let’s assume that you want to set the dipole I265DS (first dipole (D1) of S800 Spectrograph) to the rigidity (Brho) 3.7965 Tm (this works for any dipole in the S800) Let’s assume that you want to set the dipole I265DS (first dipole (D1) of S800 Spectrograph) to the rigidity (Brho) 3.7965 Tm (this works for any dipole in the S800)
Line 95: Line 124:
   * Set EPICS channel **I265DS_MAG.BSET** to 13.5454 (note, the default value of this channel is always an (impossible) 9999.1111, which is ignored. Typing 13.5454 will cause the magnet to go to 257.577 A, and **I265DS_MAG.BSET** goes back to 9999.1111. Setting the magnet this way will get you to plus or minus half a percent of where you want to be.   * Set EPICS channel **I265DS_MAG.BSET** to 13.5454 (note, the default value of this channel is always an (impossible) 9999.1111, which is ignored. Typing 13.5454 will cause the magnet to go to 257.577 A, and **I265DS_MAG.BSET** goes back to 9999.1111. Setting the magnet this way will get you to plus or minus half a percent of where you want to be.
   * At that point, the quickest way may be to use the FINE knob in the NMR controller to find the actual field value. Just turn the knob until the resonance is centered in the oscilloscope, and then read off the field value in the front panel of the controller.   * At that point, the quickest way may be to use the FINE knob in the NMR controller to find the actual field value. Just turn the knob until the resonance is centered in the oscilloscope, and then read off the field value in the front panel of the controller.
-  * Once you know the value of the field, type it (in Tesla) into the EPICS channel **I265DS_MAG.FELD**. Note that is what Daniel's program does whenever it finds  the resonance. +  * Once you know the value of the field, type it (in Tesla) into the EPICS channel **I265DS_MAG.FELD**. Note that this is what Daniel'NMR GUI program does whenever it finds  the resonance. 
   * Having done that, you can use the BARNEY 'match' button to match the magnet.   * Having done that, you can use the BARNEY 'match' button to match the magnet.
   * Repeat until satisfied.   * Repeat until satisfied.
nmr_probes.txt · Last modified: 2024/06/01 16:35 by swartzj