===== Running the S800 DAQ ===== | ===== Running the S800 DAQ ===== | ||
- | This page gives user-level instructions on how to run the S800 data acquisition (DAQ) system for an experiment. Presently this is run from the linux machine [[Software#u6pc5 (data U6)|u6pc5]] in data-U6. | + | This page gives user-level instructions on how to run the S800 data acquisition (DAQ) system for an experiment. Presently this is run from the linux machine [[Software#u4pc8 (data U4)|u4pc8]] in data-U4. |
==== Readout GUI ==== | ==== Readout GUI ==== | ||
- | The S800 Readout GUI is invoked by either clicking the icon **S800 DAQ** in the desktop of [[Software#u6pc5 (data U6)|u6pc5]] or by navigating in a Linux terminal to the directory **/ | + | The S800 Readout GUI is invoked by either clicking the icon **S800 DAQ** in the desktop of [[Software#u4pc8 (data U6)|u4pc8]] or by navigating in a Linux terminal to the directory **/ |
- | Before beginning | + | Before beginning data taking, it is necessary to initialize the system. This is done by clicking '' |
- | After initializing the contollers, the ReadoutGUI window will show the '' | + | After initializing the controllers, the ReadoutGUI window will show the '' |
Before beginning a run, you have to make sure that the ULM trigger module is properly configured. This can be done by clicking the button [[#Trigger GUI|Launch ULM GUI]] in the ReadoutGUI window. | Before beginning a run, you have to make sure that the ULM trigger module is properly configured. This can be done by clicking the button [[#Trigger GUI|Launch ULM GUI]] in the ReadoutGUI window. | ||
Line 132: | Line 132: | ||
* [[#CCUSB and/or VMUSB claimed by existing programs]] | * [[#CCUSB and/or VMUSB claimed by existing programs]] | ||
* [[#Manual rebooting]] | * [[#Manual rebooting]] | ||
+ | * [[#For hodoscope runs, cannot read signal from hodoscope]] | ||
Line 164: | Line 165: | ||
==== Manual rebooting ==== | ==== Manual rebooting ==== | ||
In the rare case that the RunControl GUI gets " | In the rare case that the RunControl GUI gets " | ||
+ | |||
+ | ==== For hodoscope runs, cannot read signal from hodoscope ==== | ||
+ | If you see this kind error message when starting the DAQ with the hodoscope in the trigger (Ext. 1 in ULM gui), check the voltages on the crate which holds the hodoscope shaper modules. If they are not at the correct values, follow this procedure: | ||
+ | - Turn crate off | ||
+ | - Unplug both shapers (for both hodoscopes). Just pull them but don’t fully remove the cards. | ||
+ | - With shapers unplugged, turn crate on and check voltages | ||
+ | - With crate on, firmly push one of the shapers in. | ||
+ | - Push second shaper in | ||
+ | - Verify crate voltages | ||
+ | This procedure works for the electronic configuration with CAEN shaper modules. | ||
+ | |||
+ | |||