The S800 [1] is a superconducting spectrograph used for reaction studies with high-energy radioactive beams produced at the NSCL Coupled-Cyclotron Facility (CCF) and the A1900 Separator [2]. It was designed for high-precision measurements of scattering angles (ΔΘ=2 msr) and momentum (p/ | The S800 [1] is a superconducting spectrograph used for reaction studies with high-energy radioactive beams produced at the NSCL Coupled-Cyclotron Facility (CCF) and the A1900 Separator [2]. It was designed for high-precision measurements of scattering angles (ΔΘ=2 msr) and momentum (p/ | ||
+ | |||
{{: | {{: | ||
- | ==== Analysis Line ==== | + | === Analysis Line === |
The analysis line extends from the object position to the target station, with a total length of 22 m. It includes four 22.5° dipoles, five quadrupole triplets, and two vertically steering magnets, assembled in two segments with configurations QQQ-H-DD-QQQ (segment 6) and QQQ-DD-H-QQQ-QQQ (segment 7) symmetrically oriented around an intermediate image plane. The maximum rigidity is 5 Tm, although it depends on the tune of the quadrupoles. The acceptances of the analysis line depends on the optical mode. | The analysis line extends from the object position to the target station, with a total length of 22 m. It includes four 22.5° dipoles, five quadrupole triplets, and two vertically steering magnets, assembled in two segments with configurations QQQ-H-DD-QQQ (segment 6) and QQQ-DD-H-QQQ-QQQ (segment 7) symmetrically oriented around an intermediate image plane. The maximum rigidity is 5 Tm, although it depends on the tune of the quadrupoles. The acceptances of the analysis line depends on the optical mode. | ||
- | ==== Spectrograph | + | === Spectrograph === |
The spectrograph consist of two quadrupoles, | The spectrograph consist of two quadrupoles, | ||
- | | Momentum Resolution (p/ | + | ^ Momentum Resolution (p/ |
- | ^ Momentum Acceptance | + | ^ Momentum Acceptance |
- | ^ Angle Resolution | + | ^ Angle Resolution |
- | ^ Solid Angle Acceptance | + | ^ Solid Angle Acceptance |
- | ^ Momentum Dispersion (x/ | + | ^ Momentum Dispersion (x/ |
- | ^ Angle Dispersion (y/b) | no colspan this time | | + | ^ Angle Dispersion (y/b) | 0.9 mm/ |
- | ^ Magnification(x/ | + | ^ Magnification(x/ |
- | ^ Focal Plane Size (x × y) | no colspan this time | | + | ^ Focal Plane Size (x × y) | 55 cm ×15 cm | |
- | ^ Maximum Rigidity | + | ^ Maximum Rigidity |
- | ^ Detector Position Resolution (x)| no colspan this time | | + | ^ Detector Position Resolution (x)| 0.3 mm | |
+ | ^ Detector Position Resolution (y)| 0.3 mm | | ||
+ | |||
+ | |||