 The main purpose of the trigger is to implement a coincidence between the S800 focal-plane fast timing scintillator [[Detectors#Plastic scintillators|E1]] and a secondary detector principally located at the target location. This coincidence is often a mandatory requirement when the trigger rate of the S800 alone (S800 singles) is too high for the data acquisition and the resulting dead time is prohibitive.  The main purpose of the trigger is to implement a coincidence between the S800 focal-plane fast timing scintillator [[Detectors#Plastic scintillators|E1]] and a secondary detector principally located at the target location. This coincidence is often a mandatory requirement when the trigger rate of the S800 alone (S800 singles) is too high for the data acquisition and the resulting dead time is prohibitive. 
  
Line 10: Line 9:
   - Direct visualization of the trigger logic and configuration    - Direct visualization of the trigger logic and configuration 
  
-Probably the most appealing feature of this module is the possibility to remotely control the trigger timing and configuration while beam is present in the vault, since most experiments are nowadays locating their electronics close to the detectors. +Probably the most appealing feature of this module is the possibility to remotely control the trigger timing and configuration while beam is present in the vault, since most experiments are nowadays locating their electronics close to the detectors. For more details on this detector check section [[Trigger#Trigger module|Trigger module]].
  
-The S800 trigger from the [[Detectors#Plastic scintillators|E1 scintillator]] provides also the reference time for the [[Detectors#Cathode Readout Drift Chambers (CRDC)|CRDCs]] as well as time-of-flight measurements. Note that because the FPGA uses a 40 MHz internal clock, the time reference of the signals in the trigger circuit are set the phase of that clock, and therefore jitter by 25 ns with respect to the source signals. This jitter is measured with a TDC and can be subtracted to the time measurements to recover the timing relative to the source signals.  
  
  
-===== Trigger Schematic =====+The S800 trigger from the [[Detectors#Plastic scintillators|E1 scintillator]] provides also the reference time for the [[Detectors#Cathode Readout Drift Chambers (CRDC)|CRDCs]] as well as time-of-flight measurements. Note that because the FPGA uses a 40 MHz internal clock, the time reference of the signals in the trigger circuit set the phase of that clock, and therefore jitter by 25 ns with respect to the source signals. This jitter is measured with a TDC and can be subtracted to the time measurements to recover the timing relative to the source signals.  
 + 
 + 
 +===== Trigger schematic ===== 
 + 
 The trigger schematic is shown on the Graphical User Interface (GUI) displayed in the figure below.  The trigger schematic is shown on the Graphical User Interface (GUI) displayed in the figure below. 
  
Line 23: Line 25:
  
  
-==== Trigger Box ====+==== Trigger box ====
 In addition to the singles and coincidence triggers, two separate trigger sources labeled "External 1" and "External 2" can be used. The various sources can be selected from the trigger box to define the raw trigger, which is then sent to a third AND gate for computer busy rejection. The busy latch (on the middle) is set by the raw trigger after a 50 ns delay, and prevents subsequent events to be accepted. It is reset by the computer once the current event has been processed. The live trigger signal feeds several gate generators which provide appropriate gates for the ADCs, QDCs, TDCs and an eventual coincidence register. Note that the trigger box contains its own coincidence register for which the input signals are delayed by 50 ns, and the gate width is set by the coincidence gate generator. The request event signal is latched before being sent to the computer.  In addition to the singles and coincidence triggers, two separate trigger sources labeled "External 1" and "External 2" can be used. The various sources can be selected from the trigger box to define the raw trigger, which is then sent to a third AND gate for computer busy rejection. The busy latch (on the middle) is set by the raw trigger after a 50 ns delay, and prevents subsequent events to be accepted. It is reset by the computer once the current event has been processed. The live trigger signal feeds several gate generators which provide appropriate gates for the ADCs, QDCs, TDCs and an eventual coincidence register. Note that the trigger box contains its own coincidence register for which the input signals are delayed by 50 ns, and the gate width is set by the coincidence gate generator. The request event signal is latched before being sent to the computer. 
  
-Since more than one trigger source can be selected, it is possible that more than one pulse is generated at the output of the trigger box, depending on the timing and shape of the source signals. When both downscaled singles and coincidences are selected for instance, the "Raw trigger" output of the trigger box may generate two pulses for a single event. A scaler connected to the "Raw trigger" output will therefore not reflect the true number of events. For this reason scalers are also connected to individual inputs of the trigger box (for more details, check section [[Electronics#Scalers]]). A trigger register word (bit pattern) is written at each occurrence of a live trigger signal. This word is the first being read out from the trigger module, prior to the time stamp ([[check trigger packet tag 0x2367 in the DAQ section for more information]]). +Since more than one trigger source can be selected, it is possible that more than one pulse is generated at the output of the trigger box, depending on the timing and shape of the source signals. When both downscaled singles and coincidences are selected for instance, the "Raw trigger" output of the trigger box may generate two pulses for a single event. A scaler connected to the "Raw trigger" output will therefore not reflect the true number of events. For this reason scalers are also connected to individual inputs of the trigger box (for more details, check the section [[Trigger#Scalers and dead time|Scalers and dead time]]). A trigger register word (bit pattern) is written at each occurrence of a live trigger signal. This word is the first being read out from the trigger module, prior to the time stamp ([[check trigger packet tag 0x2367 in the DAQ section for more information]]). 
  
-More details about the trigger box module and its FPGA schematics can be found [[here]].+More details about the trigger box module and its FPGA schematics can be found in section [[Trigger#Trigger module|Trigger module]].
  
  
Line 41: Line 43:
 ==== Busy Circuit ==== ==== Busy Circuit ====
 As each controller and external data acquisition perform their readout sequence in parallel, they have different busy times. The busy circuit following the generation of raw triggers is mainly composed of a latch that is set by the "Raw trigger" signal, and cleared by the falling edge of the OR of the individual busy signals (Busy inputs) coming from individual crate controllers or other data acquisition systems. This way the "Live trigger" signal stays true as long as the longest busy signal. The "Live trigger" signal is therefore a "global busy" signal as well. In addition, the same busy signals of individual crate controllers or other data acquisition system are used to veto the "Raw trigger" signal and prevent the generation of a live trigger. This takes care of situations where separate triggers are generated for some of the crate controllers or other data acquisition systems (such as scaler readout sequences), during which no event readout sequence should be started.  As each controller and external data acquisition perform their readout sequence in parallel, they have different busy times. The busy circuit following the generation of raw triggers is mainly composed of a latch that is set by the "Raw trigger" signal, and cleared by the falling edge of the OR of the individual busy signals (Busy inputs) coming from individual crate controllers or other data acquisition systems. This way the "Live trigger" signal stays true as long as the longest busy signal. The "Live trigger" signal is therefore a "global busy" signal as well. In addition, the same busy signals of individual crate controllers or other data acquisition system are used to veto the "Raw trigger" signal and prevent the generation of a live trigger. This takes care of situations where separate triggers are generated for some of the crate controllers or other data acquisition systems (such as scaler readout sequences), during which no event readout sequence should be started. 
- 
  
 ==== Time Stamping Scheme ==== ==== Time Stamping Scheme ====
Line 47: Line 48:
  
  
-===== Trigger module =====+====== Trigger module ======
 The S800 trigger logic is built in a LeCroy ULM2367 FPGA module. Note that this module could be replaced in the future by another FPGA module provided it has enough NIM or ECL input/outputs (such as the VME XLM72 module for instance). This section describes the functionality of the S800 trigger module and the commands used to control its parameters.  The S800 trigger logic is built in a LeCroy ULM2367 FPGA module. Note that this module could be replaced in the future by another FPGA module provided it has enough NIM or ECL input/outputs (such as the VME XLM72 module for instance). This section describes the functionality of the S800 trigger module and the commands used to control its parameters. 
  
Line 120: Line 121:
  
  
-===== Time Stamping =====+====== Time Stamping ======
 The S800 trigger provides a vetoed 10 MHz clock signal (derived from the 40 MHz FPGA clock) used for time stamping. An external clock can also be used, after selecting the appropriate check box in the [[Trigger#Trigger Schematic|GUI]]). The clock is inhibited by a "Go" signal controlled by the trigger module. While "Go" is false, all time stamp counters can be reset via CAMAC command, typically during the begin sequences of the controllers or data acquisitions (see section on [[begin sequence]]). The clock signal is released when the "Go" signal is set to true at the end of the begin sequence. This simple scheme insures that all time stamp counters are synchronized.  The S800 trigger provides a vetoed 10 MHz clock signal (derived from the 40 MHz FPGA clock) used for time stamping. An external clock can also be used, after selecting the appropriate check box in the [[Trigger#Trigger Schematic|GUI]]). The clock is inhibited by a "Go" signal controlled by the trigger module. While "Go" is false, all time stamp counters can be reset via CAMAC command, typically during the begin sequences of the controllers or data acquisitions (see section on [[begin sequence]]). The clock signal is released when the "Go" signal is set to true at the end of the begin sequence. This simple scheme insures that all time stamp counters are synchronized. 
  
Line 143: Line 144:
  
  
-===== Configuration for S800 in tandem with other detectors =====+====== Configuration for S800 in tandem with other detectors ======
 In its standard configuration, the S800 data acquisition uses one CAMAC crate and one VME crate only. The CC-USB and VM-USB crate controller modules performing the readout are connected to the latches number 1 and 2 of the trigger module, respectively. Each crate controller is configured to output their busy and end-of-event signals on their available NIM outputs, which are then connected to the appropriate inputs on the trigger module.  In its standard configuration, the S800 data acquisition uses one CAMAC crate and one VME crate only. The CC-USB and VM-USB crate controller modules performing the readout are connected to the latches number 1 and 2 of the trigger module, respectively. Each crate controller is configured to output their busy and end-of-event signals on their available NIM outputs, which are then connected to the appropriate inputs on the trigger module. 
 To incorporate an external detector in the S800 trigger logic, the same busy and end-of-event signals are required from its data acquisition system. This is to ensure that no live trigger signal is generated when any of the partners is busy or still processing an event. The 5 signals necessary between the S800 trigger and an external data acquisition system are the following:  To incorporate an external detector in the S800 trigger logic, the same busy and end-of-event signals are required from its data acquisition system. This is to ensure that no live trigger signal is generated when any of the partners is busy or still processing an event. The 5 signals necessary between the S800 trigger and an external data acquisition system are the following: 
Line 178: Line 179:
  
  
-===== Begin sequence =====+====== Begin sequence ======
 The internal "Go" state of the trigger module is controlled via CAMAC commands. When "Go" is false, the trigger and time stamp clock signals are vetoed and therefore absent. This way all time stamp counters can be safely zeroed during the beginning sequence of the data acquisition systems. The last command of the CAMAC beginning sequence sets the "Go" state to true, at which point both trigger and time stamp signals are released. This mechanism ensures that all time stamp counters are synchronized.  The internal "Go" state of the trigger module is controlled via CAMAC commands. When "Go" is false, the trigger and time stamp clock signals are vetoed and therefore absent. This way all time stamp counters can be safely zeroed during the beginning sequence of the data acquisition systems. The last command of the CAMAC beginning sequence sets the "Go" state to true, at which point both trigger and time stamp signals are released. This mechanism ensures that all time stamp counters are synchronized. 
  
Line 189: Line 190:
 The last step of the begin sequence allows enough time for the CCUSB crate controller to switch from its interactive mode to data acquisition mode. The end sequence script executed at the end of a run sets the "Go" state of the module back to false.  The last step of the begin sequence allows enough time for the CCUSB crate controller to switch from its interactive mode to data acquisition mode. The end sequence script executed at the end of a run sets the "Go" state of the module back to false. 
  
-===== Scalers and dead time ===== +====== Scalers and dead time ====== 
-The "D" connector of the trigger module is directly connected to 16 inputs of a scaler module (see mapping in the inputs and outputs section below). Scalers are connected to each of the trigger source inputs, as well as trigger box inputs. These scalers can be used to recover the number of trigger signals occurring on each of the source and trigger box inputs, in addition to the information coded for each event in the trigger register. +The "D" connector of the trigger module is directly connected to 16 inputs of a scaler module (see mapping of the inputs and outputs of the trigger module in section [[Trigger#Inputs and outputs|Inputs and outputs]]). Scalers are connected to each of the trigger source inputs, as well as trigger box inputs. These scalers can be used to recover the number of trigger signals occurring on each of the source and trigger box inputs, in addition to the information coded for each event in the trigger register. 
  
 In addition, scalers are connected to the raw and live trigger signals. For the determination of the dead time, both a free running and vetoed 10 kHz pulser signal are also connected to scalers. This is the preferred method because the pulser is not subject to possible double triggering effects like the raw trigger.  In addition, scalers are connected to the raw and live trigger signals. For the determination of the dead time, both a free running and vetoed 10 kHz pulser signal are also connected to scalers. This is the preferred method because the pulser is not subject to possible double triggering effects like the raw trigger. 