User Tools

Site Tools


tuning_the_s800_xdt

Differences

This shows you the differences between two versions of the page.

Link to this comparison view

Both sides previous revision Previous revision
Next revision
Previous revision
Next revision Both sides next revision
tuning_the_s800_xdt [2015/10/20 15:01]
pereira
tuning_the_s800_xdt [2015/10/21 16:50]
pereira [Unreacted beam]
Line 12: Line 12:
  
 === Send beam to FP === === Send beam to FP ===
 +  * Ensure that the S800 spectrograph magnets are tuned to the right rigidity
 +
 +  * Verify that the beam blocker (labeled I255 Slits) in the S3 page of Barney is open:
 +      * Expected "open" values for top and bottom slits are CT ~6.8 and CB ~3.2, respectively 
  
   * Ensure that CRAD04 (typically connected to object scintillator) is enabled with a rate limit of **20 kHz** (CRAD04 looks at E1 up FP scintillator)   * Ensure that CRAD04 (typically connected to object scintillator) is enabled with a rate limit of **20 kHz** (CRAD04 looks at E1 up FP scintillator)
  
-  * FP rate limit**6 kHz**+  * Remember: S800 FP rate limit is **6 kHz**
  
   * **[[hv bias|Bias]]** S800 FP scintillator photomultipliers   * **[[hv bias|Bias]]** S800 FP scintillator photomultipliers
       * Set bias to best guess based on previous experience for the fragment Z being used (bias will be adjusted later during tuning)       * Set bias to best guess based on previous experience for the fragment Z being used (bias will be adjusted later during tuning)
-      * Typical values:+      * Typical values from previous experiments:
           * For He-3 @ ~130 MeV/u: UP (1790 V); DOWN (1760 V)           * For He-3 @ ~130 MeV/u: UP (1790 V); DOWN (1760 V)
           * For Be-12 @ ~30 MeV/u: UP (1770 V); DOWN (1700 V)           * For Be-12 @ ~30 MeV/u: UP (1770 V); DOWN (1700 V)
Line 29: Line 33:
  
   * Remove stops to look for beam at S800 FP with **[[s800 daq tools#scaler display|scalers]]** and adjust beam rate with attenuators   * Remove stops to look for beam at S800 FP with **[[s800 daq tools#scaler display|scalers]]** and adjust beam rate with attenuators
-      * Look at FP scintillator scalers (E1 up, E1 down): +      * Look at FP scintillator scalers (E1 up, E1 down) 
-          * There are typically a few scaler counts without beam +      * There are typically a few scaler counts without beam 
-          * Ion chamber does not have scalers+
  
  
Line 40: Line 44:
  
   * Use **[[electronics overview|scope]]** to look at signal patched out to data U6 (channel #54 in data U6 patch panel)   * Use **[[electronics overview|scope]]** to look at signal patched out to data U6 (channel #54 in data U6 patch panel)
-      * This signal is sent to the CFD in data U6 +      * This signal is sent to the CANBERRA 454 Quad CFD in data U6 
-      *  +      * One of the output from this CFD is sent (via patch panel #62) to the TAC and scaler (channel OBJ.Scint) in S3. The other output goes through a passive delayed, and is sent (via patch panel #67) to the Phillips TDC 
-      * Check raising time and amplitude. Good signal: few ns raising time; 400-500 mV amplitude+      * Check raising time and amplitude. Good signal: ~10 ns raising time; 400-500 mV amplitude
    
-  * Using the scope, check the CFD setting+  * Using the scope, check the CFD setting:
       * Check CFD walk inspect signal in scope by triggering scope with CFD output       * Check CFD walk inspect signal in scope by triggering scope with CFD output
-      * Ensure that CFD delay cable is ok: about 80% of raising time of the input signal    +      * Ensure that CFD delay cable is ok: about 80% of raising time of the input signal 
 +      * Adjust CFD threshold looking at scalers. The ratio of OBJ to XFP scaler rates (channels OBJ.Scint and XFP.Scint) should reflect the transmission of the cocktail beam 
  
 +  * Adjust MCFD threshold:
 +      * Open configuration file **MCFD16.tcl** in **/user/s800/operations/daq/usb/Configs** 
 +      * The OBJ signal feeding this module is not patched out to data U6
 +      * The OBJ signal from MCFD-16 module goes to the Mesytec MTDC32 module and scaler (channel OBJ.MCFD.Scint) 
 +      * Make sure that the threshold of the XFP MCFD channel is reasonable. Rates in scaler channels XFP.Scint and XFP.MCFD.Scint should be comparable
 +      * Adjust MCFD OBJ threshold looking at scalers. The ratio of OBJ to XFP scaler rates (channels OBJ.MCFD.Scint and XFP.MCFD.Scint) should reflect the transmission of the cocktail beam        
 +      * Save new threshold in configuration file **MCFD16.tcl** 
 +   
  
   * Watch for no rate change on scaler display with a bias adjustment up or down of about 50-100 V   * Watch for no rate change on scaler display with a bias adjustment up or down of about 50-100 V
Line 54: Line 66:
 === FP scintillator setup === === FP scintillator setup ===
  
-  * Set trigger to “s800 trigger” +  * Set trigger to “s800 trigger”  
-      * Ensure that the **[[s800 daq tools#trigger GUI|trigger GUI]]** application is ready. Otherwise, open it by clicking icon **[[s800 daq tools#Run Control Window|RunControl]]** in the desktop of [[Software#u6pc5 (data U6)|u6pc5]] computer +      * Ensure that the **[[s800 daq tools#trigger GUI|trigger GUI]]** application is ready. Otherwise, open it by clicking button **[[s800 daq tools#Run Control Window|Launch ULM GUI]]** in ReadoutGUI  
-      * Under trigger tab select **s800 trigger** (which is E1 up by definition)+          * Under trigger tab select **s800 trigger** (which is E1 up by definition)
           * Deselect experiment trigger           * Deselect experiment trigger
           * SAVE TO FILE           * SAVE TO FILE
-          * Stop and start **[[s800 daq tools#Run Control Window|RunControl]]** to assert new trigger condition+          * End and Begin **[[s800 daq tools#Run Control Window|ReadoutGUI]]** to assert new trigger condition
  
   * Select **[[s800 SpecTcl|Spectcl]]** window **S800_SCINT.win**   * Select **[[s800 SpecTcl|Spectcl]]** window **S800_SCINT.win**
Line 89: Line 101:
 {{:wiki:IC-raws.png?500|IC.raw spectrum.}} {{:wiki:IC-raws.png?500|IC.raw spectrum.}}
  
-        * Gains are controlled in **s800shpini.tcl** file in directory ''s800/operations/daq/usb/Configs'' (an example of the content of this file can be seen {{:wiki:s800shpini.pdf|here}}).+        * Gains are controlled in **s800shpini.tcl** file in directory **s800/operations/daq/usb/Configs** (an example of the content of this file can be seen {{:wiki:s800shpini.pdf|here}}).
             * First shaper is for ion chamber             * First shaper is for ion chamber
             * Typically, only coarse gains are used             * Typically, only coarse gains are used
-            * Stop and start **[[s800 daq tools#Run Control Window|RunControl]]** to assert new gain values+            * End and Begin **[[s800 daq tools#Run Control Window|ReadoutGUI]]** to assert new gain values
  
  
Line 101: Line 113:
           * Patched to data-U6 on labeled connector           * Patched to data-U6 on labeled connector
           * **200 – 500 mV** signals are good           * **200 – 500 mV** signals are good
 +          * CRDC1 anode is noisier (digital noise) than CRDC2 
       * Bias CRDC1 and CRDC2. Typical starting values:       * Bias CRDC1 and CRDC2. Typical starting values:
           * For He-3 @ ~130 MeV/u: CRDC1 (Anode=1120 V, Drift=1000 V); CRDC2 (Anode=1120 V; Drift=1000 V)           * For He-3 @ ~130 MeV/u: CRDC1 (Anode=1120 V, Drift=1000 V); CRDC2 (Anode=1120 V; Drift=1000 V)
Line 112: Line 125:
       * Count rate is a little higher than on scintillator due to noise or thresholds       * Count rate is a little higher than on scintillator due to noise or thresholds
  
-  * Check **[[s800 SpecTcl|Spectcl]]** window **S800_CRDCS.win** to verify the good performance of the detectors. (The spectra for each CRDC can be checked separatelly in windows **s800_CRDC1.win** and **S800_CRDC2.win**)+  * Check **[[s800 SpecTcl|Spectcl]]** window **S800_CRDCS.win** (see figure below) to verify the good performance of the detectors. (The spectra for each CRDC can be checked separatelly in windows **s800_CRDC1.win** and **S800_CRDC2.win**)
  
       * Spectra **crdc1.raws** and **crdc2.raws** (top and middle spectra in the leftmost (first) column)       * Spectra **crdc1.raws** and **crdc2.raws** (top and middle spectra in the leftmost (first) column)
Line 119: Line 132:
           * Width of beam peak is proportional to A1900 p-acceptance in focus optics           * Width of beam peak is proportional to A1900 p-acceptance in focus optics
           * Width is narrower in match optics            * Width is narrower in match optics 
-          * Adjust anode HV to bring fuzzy maximum to around 600-700 channels (the ADC for each pad saturates at 1000 ch)+          * Adjust anode HV to bring fuzzy maximum to around 600-700 channels (saturation of each pad at 1000 ch)
  
       * Spectra **crdc1.anode_crdc1.tac** and **crdc2.anode_crdc2.tac** (top and middle spectra in the second column)       * Spectra **crdc1.anode_crdc1.tac** and **crdc2.anode_crdc2.tac** (top and middle spectra in the second column)
Line 135: Line 148:
      
       * Spectra **crdc1.xg** and **crdc2.xg** (bottom spectra in first and second columns)       * Spectra **crdc1.xg** and **crdc2.xg** (bottom spectra in first and second columns)
-          * It shows the position of the beam in the dispersive direction, evaluated by calculating the "center of gravity" +          * It shows the position of the beam in the dispersive direction, evaluated by calculating the "center of gravity". The peak should be in the middle of the spectra in order to center the beam 
  
       * Spectra **crdc1.tac** and **crdc2.tac** (bottom spectra in third and fourth columns)       * Spectra **crdc1.tac** and **crdc2.tac** (bottom spectra in third and fourth columns)
-          * They correspond to the non-dispersive position of the beam in the CRDCs. The peak should be in the middle of the spectra in order to center the beam +          * They correspond to the non-dispersive position of the beam in the CRDCs. 
  
 {{:wiki:CRDCS-example.png?850|CRDCs summary spectra}} {{:wiki:CRDCS-example.png?850|CRDCs summary spectra}}
Line 145: Line 158:
  
 === Timing setup === === Timing setup ===
 +
 +rview:
 +
 +  * There are three electronic "sources" with time information for ToF calculation: Tennelec TACs, Phillips TDC, and Mesytec MTDC. Although the "start" signal in these three sources is given by E1 up, the electronic path from the detector to the module is different
 +  * MTDC:
 +      * The OBJ and XFP signals comes from the MCFD module
 +      * The OBJ signal into the MCFD comes from the detector via S3 patch panel #94 (going to the target area) 
 +      * The XFP signal into the MCFD module comes from data-U6 patch panel #70, which is connected to the CANBERRA 454 CFD XFP output in data U6
 +      * The MTDC timing signals do not require external delay adjustments because the matching window is sufficiently wide 
 +  * Tennelec TACs:
 +      * The OBJ and XFP signals do not go through the MCFD module
 +      * The OBJ stop signal comes from data-U6 patch panel #62, which is connected to the CANBERRA 454 CFD OBJ output in data U6. The corresponding CFD input is connected to data-U6 patch panel #54
 +
 +      * The XFP stop signal comes from data-U6 patch panel #70, which is connected to the CANBERRA 454 CFD OBJ output in data U6. The corresponding CFD input receives the signal via patch panel to data U1
 +
 +
 +      * 
 +      * the detector via S3 patch panel #94 (going to the target area) 
 +      * The XFP signal into the MCFD module comes from data-U6 patch panel #70, which is connected to the CANBERRA 454 CFD output in data U6
 +      * The MTDC timing signals do not require external delay adjustments because the matching window is sufficiently wide 
 +
 +
 +
 +  * Phillips TDC:
 +      * The OBJ and XFP signals do not go through the MCFD module
 +      * The OBJ stop signal comes from data-U6 patch panel #67.
 +      *  
 +      * The XFP signal into the MCFD module comes from data-U6 patch panel #70, which is connected to the CANBERRA 454 CFD output in data U6
 +
 +      * The full range of the TACs and Phillips TDC is ~400 ns
 +      * 
  
   * See [[http://groups.nscl.msu.edu/s800/Technical/Electronics/Electronics_frameset.htm]] for background information on the trigger setup   * See [[http://groups.nscl.msu.edu/s800/Technical/Electronics/Electronics_frameset.htm]] for background information on the trigger setup
tuning_the_s800_xdt.txt · Last modified: 2023/09/22 15:15 by swartzj