Line 256: | Line 256: | ||
- | |||
- | |||
- | ===== Follow-up ===== | ||
- | |||
- | * Before leaving beam with experimenters | ||
- | * Set up current trip points on Linux HV controls | ||
- | * Values used for K-48 | ||
- | * 5 for CRDC and Ion Chamber anodes and intermediate image ppacs | ||
- | * 50 for CRDC drifts | ||
- | * 80 for IC drift | ||
- | * Ensure alarms are running | ||
- | * Make sure Linux HV GUI alarms are enabled | ||
- | * Make sure threshold on isobutane level is set up (not currently connected to alarms because they give too many false alarms when communication is lost) | ||
- | * All logs are being recorded | ||
- | * There is no log file for biases controlled by Labview | ||
- | * Linux HV | ||
- | * LabView gas handling system | ||
- | * Note in logbook | ||
- | * Scintillator biases | ||
- | * IC gate biases | ||
- | * Post reference printouts for experimenters | ||
- | * HV status: a snapshot of HV GUI | ||
- | * Gas handling system status: a snapshot of LabView window | ||
- | |||
- | * Create window configuration with summing regions to make it easier for experimenters to track efficiency/ | ||
- | |||
- | * Setting up coincidences for additional reaction settings in an experiment | ||
- | * Do not need to redo coincidence settup if secondary beam does not change | ||
- | * Might need to redo coincidence setup if secondary beam changes drastically | ||
- | |||
- | * To watch during experiment | ||
- | * Look for isobutene running out – messes up data over several hours | ||
- | |||
- | * Implementing dE- or TOF-based corrections is part of EXR | ||
- | |||
- | More detail needed | ||
- | Minimum rates required for coincidence setup | ||
- | Selection of appropriate substitute reactions for coincidence setup | ||
- | How to feel comfortable that there will not be a problem with FP detector gases running out | ||
- | Starting alarms | ||
- | Starting logging | ||