This shows you the differences between two versions of the page.
Both sides previous revision Previous revision | Next revision Both sides next revision | ||
tuning_the_s800_xdt [2015/10/27 16:43] pereira [CRDCs setup] |
tuning_the_s800_xdt [2015/10/29 09:57] pereira [Checking Particle ID and rate at S800 FP] |
||
---|---|---|---|
Line 187: | Line 187: | ||
* Select SpecTcl window **S800_PID.win** in directory **/ | * Select SpecTcl window **S800_PID.win** in directory **/ | ||
* The three columns correspond to the PID determined with the **RF-FP** ToF (left), **OBJ-FP** (center), and **XFP-FP** (right) | * The three columns correspond to the PID determined with the **RF-FP** ToF (left), **OBJ-FP** (center), and **XFP-FP** (right) | ||
- | * The first (top) row corresponds to the Phillips TDC | + | * The first (top) row corresponds to PID spectra using the Phillips TDC |
- | * The second row corresponds to the MTDC with just the first hit included | + | * The second row corresponds to PID spectra using the MTDC with __just |
- | * The third row corresponds to the ORTEC TACs. Note that there is not **RF-FP TAC** | + | * The third row corresponds to PID spectra using the MTDC gated on the " |
+ | * The fourth row corresponds to PID spectra using the ORTEC TACs. Note that there is not **RF-FP TAC** | ||
* You might need to adjust the limits of the spectra to get a good resolution | * You might need to adjust the limits of the spectra to get a good resolution | ||