tuning_the_s800_xdt [2015/10/29 13:22]
pereira [Timing setup]
tuning_the_s800_xdt [2015/10/29 13:25]
pereira [Ionization Chamber setup]
Line 74: Line 74:
           * End and Begin **[[s800 daq tools#Run Control Window|ReadoutGUI]]** to assert new trigger condition           * End and Begin **[[s800 daq tools#Run Control Window|ReadoutGUI]]** to assert new trigger condition
  
-  * Select **[[s800 SpecTcl|Spectcl]]** window **S800_SCINT.win** in directory **/user/s800/operations/spectcl/Windows**+  * Select **[[s800 SpecTcl|Spectcl]]** window **S800_SCINT.win** in directory **/user/s800/operations/spectcl/s800v7/Windows**
  
   * Adjust **[[hv bias|bias]]** looking at 2D spectra **e1.deup_e1.dedown** (showing the parameters s800.fp.e1.de_down vs. s800.fp.e1.de_up) for the FP E1 scintillator   * Adjust **[[hv bias|bias]]** looking at 2D spectra **e1.deup_e1.dedown** (showing the parameters s800.fp.e1.de_down vs. s800.fp.e1.de_up) for the FP E1 scintillator
Line 94: Line 94:
  
  
-  * Select **[[s800 SpecTcl|Spectcl]]** window **S800_IC.win** in directory **/user/s800/operations/spectcl/Windows** +  * Select **[[s800 SpecTcl|Spectcl]]** window **S800_IC.win** 
  
   * Adjust pad gains   * Adjust pad gains
Line 179: Line 178:
       * Using the cursor mouse, check the lower and higher limits defining the region in the MTDC ToF spectra with the "good" ToF peak. Do it for the three ToFs: RF-FP, OBJ-FP, and XFP-FP       * Using the cursor mouse, check the lower and higher limits defining the region in the MTDC ToF spectra with the "good" ToF peak. Do it for the three ToFs: RF-FP, OBJ-FP, and XFP-FP
       * Go to the **Variables** page in SpecTcl GUI and assign the limits to the following variables:        * Go to the **Variables** page in SpecTcl GUI and assign the limits to the following variables: 
-              * s800.fp.vmetdc.mtdc_objlow and s800.fp.vmetdc.mtdc_rfhigh for RF-XFP +              * **s800.fp.vmetdc.mtdc_objlow** and **s800.fp.vmetdc.mtdc_rfhigh** for RF-XFP 
-              * s800.fp.vmetdc.mtdc_objlow and s800.fp.vmetdc.mtdc_rfhigh for OBJ-XFP +              * **s800.fp.vmetdc.mtdc_objlow** and **s800.fp.vmetdc.mtdc_rfhigh** for OBJ-XFP 
-              * s800.fp.vmetdc.mtdc_xfplow and s800.fp.vmetdc.mtdc_xfphigh for XFP-XFP+              * **s800.fp.vmetdc.mtdc_xfplow** and **s800.fp.vmetdc.mtdc_xfphigh** for XFP-XFP
       * For each ToF, SpecTcl will search the hit number that fits in the selected region. The new MTDC ToF parameters are **s800.fp.vmetdc.mtdc_rf**, **s800.fp.vmetdc.mtdc_obj**, and **s800.fp.vmetdc.mtdc_xfp**        * For each ToF, SpecTcl will search the hit number that fits in the selected region. The new MTDC ToF parameters are **s800.fp.vmetdc.mtdc_rf**, **s800.fp.vmetdc.mtdc_obj**, and **s800.fp.vmetdc.mtdc_xfp** 
  