User Tools

Site Tools


tuning_the_s800_xdt

Differences

This shows you the differences between two versions of the page.

Link to this comparison view

Both sides previous revision Previous revision
Next revision
Previous revision
Next revision Both sides next revision
tuning_the_s800_xdt [2015/10/29 13:25]
pereira [Ionization Chamber setup]
tuning_the_s800_xdt [2015/10/30 10:25]
pereira [Setting up Reaction Settings]
Line 126: Line 126:
       * Count rate is a little higher than on scintillator due to noise or thresholds       * Count rate is a little higher than on scintillator due to noise or thresholds
  
-  * Check **[[s800 SpecTcl|Spectcl]]** window **S800_CRDCS.win** in directory **/user/s800/operations/spectcl/Windows** (see figure below) to verify the good performance of the detectors. (The spectra for each CRDC can be checked separatelly in windows **s800_CRDC1.win** and **S800_CRDC2.win**)+  * Check **[[s800 SpecTcl|Spectcl]]** window **S800_CRDCS.win** (see figure below) to verify the good performance of the detectors. (The spectra for each CRDC can be checked separatelly in windows **s800_CRDC1.win** and **S800_CRDC2.win**)
  
       * Spectra **crdc1.raws** and **crdc2.raws** (top and middle spectra in the leftmost (first) column)       * Spectra **crdc1.raws** and **crdc2.raws** (top and middle spectra in the leftmost (first) column)
Line 189: Line 189:
       * Adjust the TDC delays of E1 up, down using the Delay GUI       * Adjust the TDC delays of E1 up, down using the Delay GUI
       * In principle, the TACs delays don't need to be adjusted       * In principle, the TACs delays don't need to be adjusted
 +
 +
 +  * Check the efficiencies of the Phillips TDC, TACs, and MTDC for the OBJ-FP and XFP-FP ToFs:
 +      * Make a gate on spectrum **IC.SUM** selecting the region of interest, and call it "IC"
 +      * Looking at the ToF spectra gated on "IC" compare the recorded events in the spectra with the number of events in the gate
 +      * The window file **S800_TOF_EFFICIENCY.win** includes all the spectra needed
 +
  
 ==== Checking Particle ID and rate at S800 FP ==== ==== Checking Particle ID and rate at S800 FP ====
  
-  * Select SpecTcl window **S800_PID.win** in directory **/user/s800/operations/spectcl/s800v7/Windows** +  * Select SpecTcl window **S800_PID.win** 
       * The three columns correspond to the PID determined with the **RF-FP** ToF (left), **OBJ-FP** (center), and **XFP-FP** (right)       * The three columns correspond to the PID determined with the **RF-FP** ToF (left), **OBJ-FP** (center), and **XFP-FP** (right)
       * The first (top) row corresponds to PID spectra using the Phillips TDC       * The first (top) row corresponds to PID spectra using the Phillips TDC
Line 329: Line 336:
               * What experimenters want (e.g., if they want singles, the cut has to be more restrictive to limit acquisition deadtime)               * What experimenters want (e.g., if they want singles, the cut has to be more restrictive to limit acquisition deadtime)
           * Move blocker, decrease attenuator, repeat           * Move blocker, decrease attenuator, repeat
 +
 +      * If necessary, do the ToF corrections to improve the PID resolution (instructions [[During experiments#Particle identification corrections|here]])
  
  
tuning_the_s800_xdt.txt ยท Last modified: 2023/09/22 15:15 by swartzj