tuning_the_s800_xdt [2016/03/13 16:23]
pereira [Timing setup]
tuning_the_s800_xdt [2017/05/26 16:08]
pereira [Timing setup]
Line 68: Line 68:
  
   * Set trigger to “s800 trigger”    * Set trigger to “s800 trigger” 
-      * Ensure that the **[[s800 daq tools#trigger GUI|trigger GUI]]** application is ready. Otherwise, open it by clicking button **[[s800 daq tools#Run Control Window|Launch ULM GUI]]** in ReadoutGUI +      * Ensure that the **[[s800 daq tools#trigger GUI|trigger GUI]]** application is ready. Otherwise, open it by clicking button ''Launch ULM GUI'' in [[#ReadoutGUI|ReadoutGUI]] 
           * Under trigger tab select **s800 trigger** (which is E1 up by definition)           * Under trigger tab select **s800 trigger** (which is E1 up by definition)
           * Deselect experiment trigger           * Deselect experiment trigger
Line 171: Line 171:
       * The two spectra in the fifth row corresponds to the MTDC summary spectra of OBJ-FP and XFP-FP ToFs (zoomed in). The spectra show the ToF (vertical axis) vs. hit number (horizontal axis). Note that in a unreacted-beam setting, the first hit typically provides the "good" ToF (i.e. start and stop signals come from the same event). This is not the case in a reaction setting, where the rates in the XFP and OBJ detectors are much higher than in the FP SCI        * The two spectra in the fifth row corresponds to the MTDC summary spectra of OBJ-FP and XFP-FP ToFs (zoomed in). The spectra show the ToF (vertical axis) vs. hit number (horizontal axis). Note that in a unreacted-beam setting, the first hit typically provides the "good" ToF (i.e. start and stop signals come from the same event). This is not the case in a reaction setting, where the rates in the XFP and OBJ detectors are much higher than in the FP SCI 
       * An empty ToF spectrum means that either the delays are not right (and need to be adjusted) or the spectrum range is too narrow        * An empty ToF spectrum means that either the delays are not right (and need to be adjusted) or the spectrum range is too narrow 
-      * The MTDC spectra should never be empty because the matching window is sufficiently wide (around 4000 ns)+      * The MTDC delays should never need to be adjusted because the matching window is sufficiently wide (around 4000 ns)
  
 {{:wiki:SpecTcl-e14019-run103.jpg?850|S800_ToF.win page}} {{:wiki:SpecTcl-e14019-run103.jpg?850|S800_ToF.win page}}
Line 184: Line 184:
       * For each ToF, SpecTcl will search the hit number that fits in the selected region. The new MTDC ToF parameters are **s800.fp.vmetdc.mtdc_rf**, **s800.fp.vmetdc.mtdc_obj**, and **s800.fp.vmetdc.mtdc_xfp**        * For each ToF, SpecTcl will search the hit number that fits in the selected region. The new MTDC ToF parameters are **s800.fp.vmetdc.mtdc_rf**, **s800.fp.vmetdc.mtdc_obj**, and **s800.fp.vmetdc.mtdc_xfp** 
  
-  * If necessary, adjust delays: +  * If necessary, adjust delays of Phillips TDC (in principle, the TACs delays don't need to be adjusted)
-      * Using the [[S800 DAQ tools#Trigger GUI|ULM trigger GUI]] assign TDC-start to one of the Inspect Trigger channels and trigger the scope with it +      * Using the [[S800 DAQ tools#Trigger GUI|ULM trigger GUI]] assign TDC-start to one of the Inspect Trigger channels (patch panel #55 - #58) and trigger the scope with it (be aware that the TDC-start signal seen in the scope has an additional delay of ~200 ns due to the length of the cables between the ULM module in S3 vault and dataU6 patch panel) 
-      * Select the timing signals (Delay inspect channels) E1 up, OBJ and XFP with the [[S800 DAQ tools#Delay Window|Delay GUI]] and look at them in the scope +      * Connect the XFP and/or OBJ signals going to the Phillips TDC (via cables plugged to patch #66 and #67, respectively) to the scope 
-      * Adjust the TDC delays of OBJ and XFP using the delay boxes connected to the CANBERRA CFD 454 in data U6 +      * Adjust the TDC delays of OBJ and/or XFP with respect to the TDC start, using the delay boxes connected to the CANBERRA CFD 454 in data U6 
-      * Adjust the TDC delays of E1 updown using the Delay GUI +      * Be aware that the after adjusting OBJ and XFP delays in the scope, the signal will be further delayed by ~200 ns due to the length of the cable between dataU6 patch panel and the Phillips TDC in the S3 vault. 
-      * In principle, the TACs delays don't need to be adjusted +      * Check the corresponding ToF spectra in SpecTcl to confirm that the timings are properly adjusted.
  
   * Check the efficiencies of the Phillips TDC, TACs, and MTDC for the OBJ-FP and XFP-FP ToFs:   * Check the efficiencies of the Phillips TDC, TACs, and MTDC for the OBJ-FP and XFP-FP ToFs:
Line 283: Line 282:
  
   * Set trigger to “s800 trigger”   * Set trigger to “s800 trigger”
-      * Ensure that the **[[s800 daq tools#trigger GUI|trigger GUI]]** application is ready. Otherwise, open it by clicking icon **[[s800 daq tools#Run Control Window|RunControl]]** in the desktop of [[Software#u6pc5 (data U6)|u6pc5]] computer+      * Ensure that the **[[s800 daq tools#trigger GUI|trigger GUI]]** application is ready. Otherwise, open it by clicking ''Launch ULM GUI'' on [[#Readout GUI|Readout GUI]] in the desktop of [[Software#u6pc5 (data U6)|u6pc5]] computer
       * Under trigger tab select **s800 trigger** (which is E1 up by definition)       * Under trigger tab select **s800 trigger** (which is E1 up by definition)
           * Deselect experiment trigger           * Deselect experiment trigger
           * SAVE TO FILE           * SAVE TO FILE
-          * Stop and start **[[s800 daq tools#Run Control Window|RunControl]]** to assert new trigger condition+          * Begin a new run on [[#Readout GUI|Readout GUI]]  to assert new trigger condition
  
   * Select **[[s800 SpecTcl|Spectcl]]** window **S800_DISPMATCH.win**   * Select **[[s800 SpecTcl|Spectcl]]** window **S800_DISPMATCH.win**
tuning_the_s800_xdt.txt · Last modified: 2023/09/22 15:15 by swartzj