tuning_the_s800_xdt [2017/04/08 13:43] pereira [Dispersion Matching Mode] |
tuning_the_s800_xdt [2017/05/26 16:08] pereira [Timing setup] |
||
---|---|---|---|
Line 184: | Line 184: | ||
* For each ToF, SpecTcl will search the hit number that fits in the selected region. The new MTDC ToF parameters are **s800.fp.vmetdc.mtdc_rf**, | * For each ToF, SpecTcl will search the hit number that fits in the selected region. The new MTDC ToF parameters are **s800.fp.vmetdc.mtdc_rf**, | ||
- | * If necessary, adjust delays: | + | * If necessary, adjust delays |
- | * Using the [[S800 DAQ tools# | + | * Using the [[S800 DAQ tools# |
- | * Select | + | * Connect |
- | * Adjust the TDC delays of OBJ and XFP using the delay boxes connected to the CANBERRA CFD 454 in data U6 | + | * Adjust the TDC delays of OBJ and/or XFP with respect to the TDC start, |
- | * Adjust | + | * Be aware that the after adjusting OBJ and XFP delays |
- | * In principle, | + | * Check the corresponding ToF spectra in SpecTcl |
* Check the efficiencies of the Phillips TDC, TACs, and MTDC for the OBJ-FP and XFP-FP ToFs: | * Check the efficiencies of the Phillips TDC, TACs, and MTDC for the OBJ-FP and XFP-FP ToFs: |