This shows you the differences between two versions of the page.
Both sides previous revision Previous revision Next revision | Previous revision Next revision Both sides next revision | ||
tuning_the_s800_xdt [2017/05/26 21:58] pereira [CRDCs setup] |
tuning_the_s800_xdt [2017/06/17 14:44] pereira [Object scintillator setup] |
||
---|---|---|---|
Line 52: | Line 52: | ||
* Check CFD walk inspect signal in scope by triggering scope with CFD output | * Check CFD walk inspect signal in scope by triggering scope with CFD output | ||
* Ensure that CFD delay cable is ok: about 80% of raising time of the input signal | * Ensure that CFD delay cable is ok: about 80% of raising time of the input signal | ||
- | * Adjust CFD threshold looking at scalers. The ratio of OBJ to XFP scaler rates (channels OBJ.Scint and XFP.Scint) should reflect the transmission of the cocktail beam | + | * Adjust CFD threshold looking at scalers. |
+ | * With beam on/off, check amplitude of signals from OBJ. You should be able to clearly see the difference between noise signals and fragment-beam signals. | ||
+ | * Raise thresholds to get rid of noise signals. | ||
+ | * NOTE: Be aware that sometimes, after running for a while, the OBJ box is activated. This results in a non-negligible count rate in OBJ scalers with beam off, which comes from HIGH amplitude signals (not noise). DO NOT try to eliminate this " | ||
+ | * The ratio of OBJ to XFP scaler rates (channels OBJ.Scint and XFP.Scint) should reflect the transmission of the cocktail beam (between 60% to 90%, depending on quality of tunning) | ||
* Adjust MCFD threshold: | * Adjust MCFD threshold: | ||
Line 66: | Line 70: | ||
==== FP scintillator setup ==== | ==== FP scintillator setup ==== | ||
+ | |||
+ | * Make sure that S800 DAQ is in [[s800 daq tools# | ||
* Set trigger to “s800 trigger” | * Set trigger to “s800 trigger” |