This shows you the differences between two versions of the page.
Both sides previous revision Previous revision Next revision | Previous revision Next revision Both sides next revision | ||
tuning_the_s800_xdt [2023/09/20 13:51] swartzj |
tuning_the_s800_xdt [2023/09/22 14:48] swartzj [FP scintillator setup] |
||
---|---|---|---|
Line 44: | Line 44: | ||
* Use oscilloscope to look at analog signal patched out to data U4 | * Use oscilloscope to look at analog signal patched out to data U4 | ||
- | * Check raising | + | * Check rise time and amplitude. Good signal: ~10 ns rise time; 400-500 mV amplitude |
* Check if there are reflections (typically seen at ~300 ns after main peak) | * Check if there are reflections (typically seen at ~300 ns after main peak) | ||
Line 91: | Line 91: | ||
* Adjust biases so that unreacted beam are at 1/3 to 1/4 of dynamic range | * Adjust biases so that unreacted beam are at 1/3 to 1/4 of dynamic range | ||
* Reaction product will typically be similar enough to unreacted beam particles | * Reaction product will typically be similar enough to unreacted beam particles | ||
- | * Different particles with different energy | + | * Different particles with different energy |
| | ||
{{: | {{: |