vacuum_control [2016/11/09 14:57] pereira [Target Chamber] |
vacuum_control [2016/11/09 15:29] pereira [Venting analysis-line lower-dipole section] |
||
---|---|---|---|
Line 51: | Line 51: | ||
- Verify that the target area is vented. | - Verify that the target area is vented. | ||
- | - Ensure | + | - Using the EMIS controllers in the middle level of the S3 vault (see figure above), ensure |
- Ensure that gate valves I246GV (between lower-dipole section and target area) and I215GV (between intermediate image area and the lower-dipole section) are closed. | - Ensure that gate valves I246GV (between lower-dipole section and target area) and I215GV (between intermediate image area and the lower-dipole section) are closed. | ||
- Open S800vac with PanelMate. | - Open S800vac with PanelMate. | ||
- | - Go to page 4 (" | + | - Go to page 4 (" |
- Window "VENT STATUS" | - Window "VENT STATUS" | ||
- Click on "VENT STATUS" | - Click on "VENT STATUS" | ||
Line 62: | Line 62: | ||
+ | {{: | ||
+ | |||
+ | |||
+ | {{: | ||
===== Spectrograph ===== | ===== Spectrograph ===== | ||
This system is pumped with cryopumps. Please do not vent this section if the target chamber is under vacuum pressure. The pressure gradient between both sections would " | This system is pumped with cryopumps. Please do not vent this section if the target chamber is under vacuum pressure. The pressure gradient between both sections would " |