User Tools

Site Tools


venting_pumping_intermediate_image_box

Differences

This shows you the differences between two versions of the page.

Link to this comparison view

Both sides previous revision Previous revision
Next revision
Previous revision
Next revision Both sides next revision
venting_pumping_intermediate_image_box [2017/06/12 14:52]
pereira [Vent focal-plane chamber]
venting_pumping_intermediate_image_box [2017/06/12 14:54]
pereira [Vent focal-plane chamber]
Line 24: Line 24:
         * Box "VENT STATUS" should change from "VENT PROCEDURE STOPPED" to "WAITING FOR VENT TIMER"         * Box "VENT STATUS" should change from "VENT PROCEDURE STOPPED" to "WAITING FOR VENT TIMER"
         * Turbo gate valve **I214GV** will be automatically closed and pressure in chamber will start rising (as seen in **I215PG**) {{:wiki:iis800vac_startventing.jpg?350|S800vac Panel Mate: start venting}}.          * Turbo gate valve **I214GV** will be automatically closed and pressure in chamber will start rising (as seen in **I215PG**) {{:wiki:iis800vac_startventing.jpg?350|S800vac Panel Mate: start venting}}. 
-        * When pressure in **I215PG** reaches atmospheric pressure, stop the venting process by selecting the box "VENT STATUS" and clicking "STOP VENTING PROCESS" {{:wiki:iis800vac_stopventing.jpg?350|S800vac Panel Mate: stop venting}}. +        * When **I215PG** reads atmospheric pressure, select the box "VENT STATUS" and click "STOP VENTING PROCESS"{{:wiki:iis800vac_stopventing.jpg?350|S800vac Panel Mate: stop venting}}. 
  
         * The status in the box "VENT STATUS" should change from "WAITING FOR VENT TIMER" to "VENT PROCEDURE STOPPED"         * The status in the box "VENT STATUS" should change from "WAITING FOR VENT TIMER" to "VENT PROCEDURE STOPPED"
         * You can now open the chamber         * You can now open the chamber
  
venting_pumping_intermediate_image_box.txt ยท Last modified: 2024/01/12 09:37 by noji