Line 41: Line 41:
   * Verify that Chamber is closed and pressure-relief valve (north side of the chamber) is properly closed (NOTE: sometimes, due to overpressure from dry N2 during the venting procedure, the O-ring in pressure-relief valve flange is moved out of place)   * Verify that Chamber is closed and pressure-relief valve (north side of the chamber) is properly closed (NOTE: sometimes, due to overpressure from dry N2 during the venting procedure, the O-ring in pressure-relief valve flange is moved out of place)
   * Verify that PPACs are not biased by opening the [[HV bias|HV GUI]]   * Verify that PPACs are not biased by opening the [[HV bias|HV GUI]]
-  * IMPORTANT: Close manual roughing pump valve close to the Intermediate-image chamber +  * IMPORTANT: Close manual roughing pump valve (close to the Intermediate-image chamber) {{:wiki:manualvalverough.jpg?500|Manual valve rough pump}}.  
   * Insert PPACs/turn on light (Diagnostic Drives PanelMate) and verify condition through sight window on south side of the chamber (be careful no to bump the bullet camera).   * Insert PPACs/turn on light (Diagnostic Drives PanelMate) and verify condition through sight window on south side of the chamber (be careful no to bump the bullet camera).
   * Open **S800Vac** and go to page 03 ("Intermediate image box vacuum system sec 15")   * Open **S800Vac** and go to page 03 ("Intermediate image box vacuum system sec 15")
Line 56: Line 57:
   * When the pressure ready by **I215PG** is few mTorr, you can turn on the ion gauge **I213IG**. Good vacuum will take ~1-2 hours.   * When the pressure ready by **I215PG** is few mTorr, you can turn on the ion gauge **I213IG**. Good vacuum will take ~1-2 hours.
   * NOTE: WHAT TO DO IF PUMPDOWN TIMER EXPIRES BEFORE TURBO PUMP IS ON   * NOTE: WHAT TO DO IF PUMPDOWN TIMER EXPIRES BEFORE TURBO PUMP IS ON
 +  * At the end of the procedure, close the manual roughing pump valve 
  