 ===== Filling Intermediate-image delay PPAC ===== ===== Filling Intermediate-image delay PPAC =====
 +Requirements: PanelMate files **S3_Gas_Handling** and **S800Vac**
  
 +  * Verify that PPACs are not biased: 
 +        * Check that the detectors are unbiased by opening the [[HV bias|HV GUI]] 
 +        * Notify the system of the bias-off state by selecting the "I214BON_CONFIRM" box in **S3_Gas_Handling** and clicking "PP BIAS OFF CONFIRMED" 
 +  * Verify that PPACs are in automatic mode with gas removed (**S3_Gas_Handling**) 
 +        * The "GH System Status box should read "Auto Mode Gas Removed" 
 +        * If gas is removed but the system is not in the automatic mode, consult the Beam Physicists or Device Physicist. This typically requires to verify that the system is in a controlled status. In particular, the "I214BON_CONFIRM" box should indicate "PP BIAS OFF CONFIRMED", and the only that should be open in **S3_Gas_Handling** are the "Det Bypass" (V5) and "Chmbr Bypass" (V6)   
 +  * Confirm that PanelMate box "I214P_AUTO/MAN Pre Set" reads 5 Torr (the set gas pressure in the detector)  
 +  * PanelMate box "Gas Flow Control" should indicate "OK to Start Gas Flow". Select that box and click on "Start Gas Flow" 
 +        *  
 +  *  
 +  * Open **S800Vac** and go to page 03 ("Intermediate image box vacuum system sec 15") 
 +  * Isolate chamber by closing gate valves **I212GV** and **I215GV** 
 +  * [[Vacuum control|Turn off]] ion gauge **I213IG**  
 +  * Check box "PUMPDOWN STATUS INTERMEDIATE IMAGE BOX". It should read "Good vacuum" 