User Tools

Site Tools


venting_pumping_intermediate_image_box

Differences

This shows you the differences between two versions of the page.

Link to this comparison view

Both sides previous revision Previous revision
Next revision
Previous revision
Next revision Both sides next revision
venting_pumping_intermediate_image_box [2018/12/19 11:39]
pereira [Filling Intermediate-image delay PPAC]
venting_pumping_intermediate_image_box [2018/12/19 11:43]
pereira [Filling/Removing gas in Intermediate-image delay PPAC]
Line 3: Line 3:
   * [[#Vent Intermediate-image chamber|Venting Intermediate-image chamber]]   * [[#Vent Intermediate-image chamber|Venting Intermediate-image chamber]]
   * [[#Pumping Intermediate-image chamber]]   * [[#Pumping Intermediate-image chamber]]
-  * [[#Filling Intermediate-image delay PPAC]]+  * [[#Filling/Removing gas in Intermediate-image delay PPAC]]
  
 ===== General information ===== ===== General information =====
Line 73: Line 73:
  
  
-===== Filling Intermediate-image delay PPAC =====+===== Filling/Removing gas in Intermediate-image delay PPAC ===== 
 + 
 +{{:wiki:ghsiippacs.png?1000|Gas Handling System of S800 Intermediate Image}}  
 Requirements: PanelMate files **S3_Gas_Handling** and **S800Vac** Requirements: PanelMate files **S3_Gas_Handling** and **S800Vac**
 +
 +=== Filling Intermediate-image delay PPAC with gas ====
   * Verify that PPACs are not biased:   * Verify that PPACs are not biased:
         * Check that the detectors are unbiased by opening the [[HV bias|HV GUI]]         * Check that the detectors are unbiased by opening the [[HV bias|HV GUI]]
venting_pumping_intermediate_image_box.txt · Last modified: 2024/01/12 09:37 by noji