venting_pumping_intermediate_image_box [2023/02/13 10:44]
noji Updated descriptions about the II-GHS based upon "S800 Intermediate Image PPAC Gas Handling System Operation Manual" (v4) provided to us on 2/8/2023 by B. Ewert.
venting_pumping_intermediate_image_box [2024/01/11 09:50]
noji [Emptying Detectors]
Line 1: Line 1:
-====== Venting/pumping Intermediate Image box and filling PPACs with gas ======+====== Venting/pumping II chamber & emptying/filling gaseous detectors ======
   * [[#General information]]   * [[#General information]]
-  * [[#Vent Intermediate-image chamber|Venting Intermediate-image chamber]] +  * [[#Emptying Detectors]] 
-  * [[#Pumping Intermediate-image chamber]] +  * [[#Filling and Running Detectors]] 
-  * [[#Filling/Removing gas in Intermediate-image delay PPAC]] +  * [[#Troubleshooting]]
 ===== General information ===== ===== General information =====
 The procedures to vent/pump the S800 Intermediate Image box, and/or filling the PPAC detectors with gas must be done carefully to ensure that the PPACs are not damaged. For instance, before venting the chamber, we have to make sure that the PPACs are empty and equalized with respect to the vacuum in the chamber. In order to avoid errors, an automated gas handling system (II-GHS) was designed to perform these tasks safely. The II-GHS equipment (valves, flow-meter, pipes, isobutane gas bottle. etc.) are located on the west balcony of the top level in the S3 vault (see figure below): The procedures to vent/pump the S800 Intermediate Image box, and/or filling the PPAC detectors with gas must be done carefully to ensure that the PPACs are not damaged. For instance, before venting the chamber, we have to make sure that the PPACs are empty and equalized with respect to the vacuum in the chamber. In order to avoid errors, an automated gas handling system (II-GHS) was designed to perform these tasks safely. The II-GHS equipment (valves, flow-meter, pipes, isobutane gas bottle. etc.) are located on the west balcony of the top level in the S3 vault (see figure below):
Line 30: Line 29:
     * Jorge noted on 2/12/2023 that this step did not work; he reported that when he clicked **PUMP** nothing happened. Instead, he had to “manually” enter the PV name associated with this command and changed from 0 to 1. He speculated that this was most likely due to a CS Studio problem.     * Jorge noted on 2/12/2023 that this step did not work; he reported that when he clicked **PUMP** nothing happened. Instead, he had to “manually” enter the PV name associated with this command and changed from 0 to 1. He speculated that this was most likely due to a CS Studio problem.
  
 +===== Venting Chamber =====
 ===== Filling and Running Detectors ===== ===== Filling and Running Detectors =====
   - If in Vent Mode, click **PUMP** then click **RUN** to enter Run Mode.   - If in Vent Mode, click **PUMP** then click **RUN** to enter Run Mode.
venting_pumping_intermediate_image_box.txt · Last modified: 2024/01/12 09:37 by noji