User Tools

Site Tools


venting_pumping_intermediate_image_box

Differences

This shows you the differences between two versions of the page.

Link to this comparison view

Both sides previous revision Previous revision
Next revision
Previous revision
Next revision Both sides next revision
venting_pumping_intermediate_image_box [2023/02/13 10:44]
noji Updated descriptions about the II-GHS based upon "S800 Intermediate Image PPAC Gas Handling System Operation Manual" (v4) provided to us on 2/8/2023 by B. Ewert.
venting_pumping_intermediate_image_box [2024/01/11 10:07]
noji [Venting the Chamber]
Line 1: Line 1:
-====== Venting/pumping Intermediate Image box and filling PPACs with gas ======+====== Venting/pumping II chamber & emptying/filling gaseous detectors ======
   * [[#General information]]   * [[#General information]]
-  * [[#Vent Intermediate-image chamber|Venting Intermediate-image chamber]] +  * [[#Emptying Detectors]] 
-  * [[#Pumping Intermediate-image chamber]] +  * [[#Filling and Running Detectors]] 
-  * [[#Filling/Removing gas in Intermediate-image delay PPAC]] +  * [[#Troubleshooting]]
 ===== General information ===== ===== General information =====
 The procedures to vent/pump the S800 Intermediate Image box, and/or filling the PPAC detectors with gas must be done carefully to ensure that the PPACs are not damaged. For instance, before venting the chamber, we have to make sure that the PPACs are empty and equalized with respect to the vacuum in the chamber. In order to avoid errors, an automated gas handling system (II-GHS) was designed to perform these tasks safely. The II-GHS equipment (valves, flow-meter, pipes, isobutane gas bottle. etc.) are located on the west balcony of the top level in the S3 vault (see figure below): The procedures to vent/pump the S800 Intermediate Image box, and/or filling the PPAC detectors with gas must be done carefully to ensure that the PPACs are not damaged. For instance, before venting the chamber, we have to make sure that the PPACs are empty and equalized with respect to the vacuum in the chamber. In order to avoid errors, an automated gas handling system (II-GHS) was designed to perform these tasks safely. The II-GHS equipment (valves, flow-meter, pipes, isobutane gas bottle. etc.) are located on the west balcony of the top level in the S3 vault (see figure below):
  
-{{:wiki:IIGHShardware.jpg?600|Gas Handling System of S800 Intermediate Image}} +{{:wiki:IIGHShardware.jpeg|Gas Handling System of S800 Intermediate Image}} 
  
 The GHS is operated remotely through CS studio. Its page can be found on the EXP network through FRIB EXP Main → S800 → GHS Pages (Secondary Pages) → Intermediate Image PPAC and is shown below.  The GHS is operated remotely through CS studio. Its page can be found on the EXP network through FRIB EXP Main → S800 → GHS Pages (Secondary Pages) → Intermediate Image PPAC and is shown below. 
Line 30: Line 29:
     * Jorge noted on 2/12/2023 that this step did not work; he reported that when he clicked **PUMP** nothing happened. Instead, he had to “manually” enter the PV name associated with this command and changed from 0 to 1. He speculated that this was most likely due to a CS Studio problem.     * Jorge noted on 2/12/2023 that this step did not work; he reported that when he clicked **PUMP** nothing happened. Instead, he had to “manually” enter the PV name associated with this command and changed from 0 to 1. He speculated that this was most likely due to a CS Studio problem.
  
 +===== Venting the Chamber =====
 +
 +From the S800 Main Vacuum page, go to the Intermediate Image - RBT_BTS32 D2430 Pumping Station page, and ensure  
 +  * that the gate valves named D2426 BGV and D2436 BGV are closed, and that the IG is off, and
 +  * that the turbo pump gate valve TVG is closed.
 +
 +To vent the chamber:
 +  - Monitor **RBT_BTS32:BPG_D2430:VP_RD** and **S800_GHS:PI_D2431:V_RD_AB** in data browser,
 +  - close **RV**, the valve to the roughing pump, and
 +  - open **Vent**.
 +
 +The PPAC pressure S800_GHS:PI_D2431:V_RD_AB should closely follow the chamber pressure  
 +RBT_BTS32:BPG_D2430:VP_RD
 +===== Evacuating the Chamber =====
 +
 +{{:wiki:iimanualvalve.jpeg|}}
 ===== Filling and Running Detectors ===== ===== Filling and Running Detectors =====
   - If in Vent Mode, click **PUMP** then click **RUN** to enter Run Mode.   - If in Vent Mode, click **PUMP** then click **RUN** to enter Run Mode.