venting_pumping_intermediate_image_box [2024/01/11 10:32]
noji [Emptying Detectors]
venting_pumping_intermediate_image_box [2024/01/11 10:36]
noji [Venting the Chamber]
Line 32: Line 32:
   * that the PPACs have been unbiased,   * that the PPACs have been unbiased,
   * that this pumping station has been isolated by having the gate valves **D2426 BGV** and **D2436 BGV** closed, and that the IG has been turned off, and   * that this pumping station has been isolated by having the gate valves **D2426 BGV** and **D2436 BGV** closed, and that the IG has been turned off, and
-  * that the turbo pump gate valve TVG is closed.+  * that the turbo pump gate valve **TVG** is closed
 + 
 +The TMP can be turned off at this time.
  
 To vent the chamber: To vent the chamber:
-  - Monitor **RBT_BTS32:BPG_D2430:VP_RD** and **S800_GHS:PI_D2431:V_RD_AB** in a data browser,+  - Put the GHS in vent mode by clicking **VENT**, 
 +  - monitor **RBT_BTS32:BPG_D2430:VP_RD** and **S800_GHS:PI_D2431:V_RD_AB** in a data browser,
   - close **RV**, the valve to the roughing pump, and   - close **RV**, the valve to the roughing pump, and
   - open **Vent**.   - open **Vent**.