Line 32: | Line 32: | ||
* that the PPACs have been unbiased, | * that the PPACs have been unbiased, | ||
* that this pumping station has been isolated by having the gate valves **D2426 BGV** and **D2436 BGV** closed, and that the IG has been turned off, and | * that this pumping station has been isolated by having the gate valves **D2426 BGV** and **D2436 BGV** closed, and that the IG has been turned off, and | ||
- | * that the turbo pump gate valve TVG is closed. | + | * that the turbo pump gate valve **TVG** is closed. |
+ | |||
+ | The TMP can be turned off at this time. | ||
To vent the chamber: | To vent the chamber: |